Related Experiment Video
Updated: Aug 24, 2025

A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
Computational design and experimental analysis of a novel visor for COVID-19 patients receiving high-flow nasal
Masooma Ijaz1, Sorcha Ni Fhrighil1, Rory Brett1
1Mechanical Engineering, College of Engineering and Informatics, National University of Ireland, Galway, Ireland.
Abstract:
The Covid-19 global pandemic has reshaped the requirements of healthcare sectors worldwide. Following the exposure risks associated with Covid-19, this paper aims to design, optimise, and validate a wearable medical device that reduces the risk of transmission of contagious droplets from infected patients in a hospital setting. This study specifically focuses on those receiving high-flow nasal oxygen therapy. The design process consisted of optimising the geometry of the visor to ensure that the maximum possible percentage of harmful droplets exhaled by the patient can be successfully captured by a vacuum tube attached to the visor. This has been completed by deriving a number of concept designs and assessing their effectiveness, based on numerical analysis, computational fluid dynamics (CFD) simulations and experimental testing. The CFD results are validated using various experimental methods such as Schlieren imaging, particle measurement testing and laser sheet visualisation. Droplet capturing efficiency of the visor was measured through CFD and validated through experimental particle measurement testing. The results presented a 5% deviation between CFD and experimental results. Also, the modifications based on the validated CFD results improved the visor effectiveness by 47% and 38% for breathing and coughing events, respectively.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:23Retinal Vascular Reactivity as Assessed by Optical Coherence Tomography Angiography
Published on: March 26, 2020
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Suctioning the Nasopharyngeal Airway
Equipment Required