Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
Comparison of the Acurate Neo Vs Neo2 Transcatheter Heart Valves
Won-Keun Kim1, Clemens Eckel, Matthias Renker
1Kerckhoff Heart Center, Department of Cardiology, Benekestrasse 2-8, 61231 Bad Nauheim Germany. w.kim@kerckhoff-klinik.de.
Background:
Few data exist on immediate outcomes of the next-generation Acurate neo2 prosthesis (Boston Scientific), which is distinguished by an active sealing mechanism. We sought to determine procedural outcomes of transfemoral transcatheter aortic valve replacement using the neo2 in comparison with its predecessor, the Acurate neo.
Methods:
In this retrospective analysis, consecutive neo2 and neo cases were compared from 2 high-volume centers. The primary outcome of interest was the rate of relevant paravalvular regurgitation (PVR), defined as PVR ≥ moderate, or valve-in-valve and/or surgical aortic valve replacement for PVR ≥ moderate. Secondary outcomes of interest were assessed according to Valve Academic Research Consortium (VARC)-3 criteria. Logistic regression analysis was used to identify predictors of relevant PVR.
Results:
A total of 810 neo2 and 2055 neo cases comprised the study cohort. The rate of relevant PVR was significantly lower in the neo2 group (2.7% vs 4.5%; P=.04). The technical success rate was numerically higher in the neo2 group (91.5% vs 89.3%; P=.10) and the rate of device success at 30 days was significantly higher (86.5% vs 82.9%; P=.02). In the neo group, a greater amount of aortic valve calcification (AVC), the presence of eccentric AVC, less oversizing, and a higher sinotubular junction annulus index were predictors of relevant PVR, whereas in the neo2 population only the presence of eccentric AVC, less oversizing, and a higher sinotubular junction annulus index was predictive.
Conclusion:
The neo2 valve shows superior outcomes over the neo valve, with a lower burden of PVR and a higher device success rate at 30 days.
More Related Videos
Related Concept Videos
Mitral Stenosis II: Clinical features and Diagnostic Tests
Mitral Regurgitation II: Clinical Features and Diagnostic Tests
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Cardiac Catheterization II: Right Heart Catheterization