Accelerated 3D high-resolution T2-weighted breast MRI with deep learning constrained compressed sensing, comparison
Fan Yang1, Xuelin Pan1, Ke Zhu1
1Department of Radiology, West China Hospital of Sichuan University, Chengdu, China.
Purpose:
To evaluate the feasibility of isotropic 3D high-resolution T2-weighted imaging (T2WI) MRI sequences and compare the images reconstructed by integrating artificial intelligence-compressed sensing (AI-CS), compressed sensing (CS), and conventional 2D T2WI sequences for quality.
Materials And Methods:
Fifty-two female patients (ages: 26-80 years) with suspected breast cancer were enrolled. They underwent breast MRI examinations using three sequences: conventional T2WI, CS 3D T2WI, and AI-CS 3D T2WI. Image quality, signal-to-noise ratio (SNR), contrast-to-noise ratio, tumor volume, and maximal tumor diameter were compared using the Friedman test. Image quality was scored on a 5-point scale, with 1 indicating nonassessable quality and 5 indicating excellent quality. Tumor volume and maximal tumor diameter were compared based on AI-CS 3D T2WI (slightly high signal), conventional T2WI, and dynamic contrast-enhanced (DCE) sequences.
Results:
All three T2WI were successfully performed in all patients. 3D CS and AI-CS were significantly better than conventional T2WI in terms of lesion conspicuity and morphology, structural details, overall image quality, diagnostic information for breast lesions, and breast tissue delineation (P < 0.001). The SNR of conventional T2WI was significantly higher for 3D T2WI sequences. The contrast-to-noise ratio was significantly higher for AI-CS 3D T2WI than for conventional T2WI sequence. There was no significant difference in tumor volume between DCE (8.08 ± 16.51) and AI-CS 3D T2WI (8.25 ± 16.29) sequences and no significant differences in tumor diameter among DCE, AI-CS 3D T2WI, and conventional T2WI sequences.
Conclusion:
Isotropic-resolution 3D T2WI sequences can be acquired using AI-CS while maintaining image quality and diagnostic value, which may pave the way for isotropic 3D high-resolution T2WI for clinical application.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging