Comparing the RETeval® portable ERG device with more traditional tabletop ERG systems in normal subjects and selected
Jia Yue You1, Allison L Dorfman1, Mathieu Gauvin1
1Department of Ophthalmology, Research Institute of the McGill University Health Center and the Montreal Children's Hospital, Glen Site, Block E, Room EM03238 (Program Mail Drop Point EM3211), 1001 Decarie Boulevard, Montréal, QC, H4A 3J1, Canada.
Purpose:
Our study aimed to determine if ISCEV standard-like ERGs recorded with the LKC RETeval® portable ERG unit compared to those obtained using the more traditional tabletop unit.
Methods:
ERGs recorded from normal subjects and patients affected with retinal ON and OFF pathway anomalies were compared. Analysis included peak time and amplitude measurements as well as time-frequency domain analysis with the discrete wavelet transform of waveforms obtained with the two systems.
Results:
Although both systems were similarly able to record reliable and highly reproducible ERG responses, there were major discrepancies in ERG responses between the portable and tabletop units, pointing toward a weaker stimulation of the retinal OFF pathway with the portable RETeval® unit.
Conclusion:
The portable RETeval® unit appears to be able to record highly reproducible and diagnostically useful clinical ERGs, albeit with some significant differences in waveform composition compared to those obtained with more standard tabletop systems. Given the unknown origin of these waveform discrepancies, if left uncorrected, these differences could potentially lead to erroneous interpretation when used in the clinical context and/or compared to ERGs recorded using more traditional table top units. Clearly, more research is warranted before handheld devices, such as the RETeval®, can be homologated as a diagnostically sound ERG devices.