Analysis of whole-body MRI artifacts in a pediatric population with a special emphasis on the effect of hands

Solène Le Cam1, Paul de Boissieu2, Jean-Paul Teglas3

  • 1Faculty of Medicine, Paris Saclay Université, 94270 Le Kremlin, Bicêtre, France; AP-HP, Bicêtre Hospital, Pediatric Imaging Department, 94270 Le Kremlin, Bicêtre, France.

Insights

Whole-body MRI (WB-MRI) in children reveals varying artifact prevalence between STIR and T1-weighted sequences. Optimizing patient positioning, particularly hand placement, can reduce artifacts for improved diagnostic accuracy.

Area of Science:

  • Radiology
  • Medical Imaging
  • Pediatric Imaging

Background:

  • Whole-body magnetic resonance imaging (WB-MRI) is increasingly utilized in pediatric care.
  • Understanding imaging artifacts is crucial for accurate interpretation in pediatric WB-MRI.

Purpose of the Study:

  • To determine the prevalence of artifacts in pediatric whole-body MRI (WB-MRI) examinations.
  • To identify the causes and contributing factors of these artifacts, including sequence type and patient positioning.

Main Methods:

  • Retrospective analysis of 107 pediatric WB-MRI examinations (ages 2-16).
  • Artifacts were categorized and analyzed across short-tau inversion recovery (STIR) and T1-weighted sequences.
  • Prevalence was assessed concerning patient age and hand position.

Main Results:

  • A total of 3436 artifacts were identified.
  • STIR sequences showed higher rates of 'noise' and 'cross-talk' artifacts compared to T1-weighted sequences.
  • T1-weighted sequences exhibited more 'equipment' and 'stair-step' artifacts. Specific hand positions influenced artifact occurrence, with implications for wrists and abdomen. Older children (>10 years) had more 'susceptibility' and 'equipment' artifacts.

Conclusions:

  • Artifact prevalence differs significantly between STIR and T1-weighted sequences in pediatric WB-MRI.
  • The combined use of both sequence types is recommended for comprehensive evaluation.
  • Optimizing patient hand positioning is essential for minimizing artifacts and ensuring diagnostic quality and safety.
Abstract