The influence of patient positioning on radiation dose in CT imaging: A narrative review
Yazan Al-Hayek1, Xiaoming Zheng2, Christopher Hayre3
1School of Dentistry and Medical Sciences, Faculty of Science and Health, Charles Sturt University, Wagga Wagga, NSW, 2650, Australia; Department of Medical Imaging, Faculty of Applied Health Sciences, The Hashemite University, Zarqa, 13133, Jordan.
Background And Purpose:
Although it is fundamental for optimal scanner operation, it is generally accepted that accurate patient centring cannot always be achieved. This review aimed to examine the reported knowledge of the negative impact of patient positioning on radiation dose and image quality during CT imaging. Furthermore, the study evaluated the current optimisation tools and techniques used to improve patient positioning relative to the gantry iso-center.
Methodology:
A comprehensive search through the databases PubMed, Ovid, and Google Scholar was performed. Keywords included patient off-centring, patient positioning, localiser radiograph orientation, radiation dose, and automatic patient positioning (including synonyms). The search was limited to full-text articles that were written in English. After initial title and abstract screening, a total of 52 articles were identified to address the aim of the review. No limitations were imposed on the year of publication.
Results:
Vertical off-centring was reported in up to 95% of patients undergoing chest and abdominal CT examinations, showing a significant influence on radiation dose. Depending on the scanner model and vendor, localiser orientation, bowtie filter used, and patient size, radiation dose varied from a decrease of 36% to an increase of 91%. A significant dose reduction was demonstrated when utilising an AP localiser, aligning with the trend for radiographers to off-center patients below the gantry iso-centre. Utilizing a 3D camera for body contour detection allowed for more accurate patient positioning and promoted further dose reduction.
Conclusion:
Patient positioning has shown significant effects on radiation dose and image quality in CT. Developing a good understanding of the key factors influencing patient dose (off-centring direction, localiser orientation, patient size and bowtie filter selection) is critical in optimising CT scanning practices. Utilising a 3D camera for body contour detection is strongly recommended to improve patient positioning accuracy, image quality and to minimise patient dose.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Anatomical Positions
The body is upright, facing forward, and standing erect.
The feet are parallel and flat on the floor.
The arms are hanging by the...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography