Objective:
to analyze the cleaning process of gastroscopes, colonoscopes and duodenoscopes in eight in-hospital health services.
Method:
a cross-sectional study conducted with 22 endoscopes (eight gastroscopes, eight colonoscopes and six duodenoscopes), and microbiological analysis of 60 samples of air/water channels (all endoscopes) and elevator (duodenoscopes), in addition to protein testing. Descriptive statistics with calculation of frequencies and central tendency measures was used in data analysis.
Results:
the processing of 22 endoscopes was monitored with microbiological analysis for 60 channels. In the pre-cleaning procedure, in 82.3% (14/17) of the devices, gauze was used in cleaning the insertion tube. Incomplete immersion of the endoscope in detergent solution occurred in 72.3% (17/22) of the cases, and in 63.6% (14/22) there was no standardization of filling-in of the channels. Friction of the biopsy channel was not performed in 13.6% (3/22) of the devices. In the microbiological analysis, 25% (7/32) of the samples from the stored endoscopes were positive for microbial growth (from 2x101 to 9.5x104 CFU/mL), while after processing, contamination was 32% (9/28). Protein residues in the elevator channel were detected in 33% of duodenoscopes.
Conclusion:
the results indicate important gaps in the stages of pre-cleaning and cleaning of endoscopes that, associated with presence of protein residues and growth of microorganisms of epidemiological importance, indicate limitations in safety of the processing procedures, which can compromise the disinfection processes and, consequently, their safe use among patients subjected to such tests.
More Related Videos
15:49Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
03:58Author Spotlight: Advancing Awake Nasotracheal Intubation with Flexible Video Rhino-Laryngoscopes
Published on: August 2, 2024
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures II: Colonoscopy
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...