Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Diagnostic Accuracy of a Portable ECG Device in Rowing Athletes
Fiona Wilson1, Cliodhna McHugh2, Caroline MacManus3
1Discipline of Physiotherapy, School of Medicine, Trinity College Dublin, D08 W9RT Dublin, Ireland.
Background:
Athletes can experience exercise-induced transient arrythmias during high-intensity exercise or competition, which are difficult to capture on traditional Holter monitors or replicate in clinical exercise testing. The aim of this study was to investigate the reliability of a portable single channel ECG sensor and data recorder (PluxECG) and to evaluate the confidence and reliability in interpretation of ECGs recorded using the PluxECG during remote rowing.
Methods:
This was a two-phase study on rowing athletes. Phase I assessed the accuracy and precision of heart rate (HR) using the PluxECG system compared to a reference 12-lead ECG system. Phase II evaluated the confidence and reliability in interpretation of ECGs during ergometer (ERG) and on-water (OW) rowing at moderate and high intensities. ECGs were reviewed by two expert readers for HR, rhythm, artifact and confidence in interpretation.
Results:
Findings from Phase I found that 91.9% of samples were within the 95% confidence interval for the instantaneous value of the changing exercising HR. The mean correlation coefficient across participants and tests was 0.9886 (σ = 0.0002, SD = 0.017) and between the two systems at elevated HR was 0.9676 (σ = 0.002, SD = 0.05). Findings from Phase II found significant differences for the presence of artifacts and confidence in interpretation in ECGs between readers' for both intensities and testing conditions. Interpretation of ECGs for OW rowing had a lower level of reader agreement than ERG rowing for HR, rhythm, and artifact. Using consensus data between readers' significant differences were apparent between OW and ERG rowing at high-intensity rowing for HR (p = 0.05) and artifact (p = 0.01). ECGs were deemed of moderate-low quality based on confidence in interpretation and the presence of artifacts.
Conclusions:
The PluxECG device records accurate and reliable HR but not ECG data during exercise in rowers. The quality of ECG tracing derived from the PluxECG device is moderate-low, therefore the confidence in ECG interpretation using the PluxECG device when recorded on open water is inadequate at this time.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Dysrhythmias V: Evaluating Dysrhythmias
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Holter Monitor: 24-Hour Monitoring
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....