Novel Ex Vivo DOAC Removal Methods Reduce Interference in Lupus Anticoagulant Testing
Paula Savola1, Marja Lemponen1, Lotta Joutsi-Korhonen1
1Department of Clinical Chemistry, HUS Diagnostic Center, Helsinki University Hospital and University of Helsinki, 00290 Helsinki, Finland.
Direct oral anticoagulant (DOAC) removal methods effectively eliminate drug concentrations, significantly reducing false positives in lupus anticoagulant (LAC) testing. This improves diagnostic accuracy and minimizes unnecessary follow-up tests for patients on DOAC therapy.
Area of Science:
- Clinical Chemistry
- Hematology
- Diagnostic Laboratory Science
Background:
- Direct oral anticoagulants (DOACs) are widely used but interfere with laboratory coagulation tests.
- This interference can lead to inaccurate results, particularly in lupus anticoagulant (LAC) testing.
- Accurate LAC testing is crucial for diagnosing thrombotic disorders.
Purpose of the Study:
- To evaluate the efficacy of commercial DOAC removal methods (DOAC Filter® and DOAC-Stop™) in eliminating DOAC concentrations.
- To assess the impact of these removal methods on reducing false-positive results in LAC screening.
- To determine if DOAC removal can decrease the need for further confirmatory or mixing tests in LAC diagnostics.
Main Methods:
- Fifty patient samples with high DOAC concentrations (apixaban, dabigatran, edoxaban, rivaroxaban) were analyzed.
- DOACs were removed ex vivo using either DOAC Filter® (n=28) or DOAC-Stop™ (n=22).
- Commercial control and calibrator samples were also tested (n=13 for DOAC Filter®, n=14 for DOAC-Stop™).
- LAC screening was performed before and after DOAC removal.
Main Results:
- Both DOAC Filter® and DOAC-Stop™ effectively reduced DOAC concentrations to a median of 0 ng/mL.
- Prior to DOAC removal, 96% of patient samples and over 90% of control samples showed false-positive LAC results.
- After DOAC removal, LAC screening became negative in 61% of patient samples treated with DOAC Filter® and 45% with DOAC-Stop™.
- All control samples tested negative for LAC after DOAC removal.
Conclusions:
- Ex vivo DOAC removal significantly reduces false-positive results in lupus anticoagulant screening.
- DOAC removal methods effectively eliminate drug interference, improving the reliability of LAC testing.
- These methods can halve the requirement for additional confirmatory tests, streamlining patient management and reducing healthcare costs.