Calculating Heart Rate Variability from ECG Data from Youth with Cerebral Palsy During Active Video Game Sessions
Published on: June 5, 2019
Validity of Ultra-Short-Term HRV Analysis Using PPG-A Preliminary Study
Aline Taoum1, Alexis Bisiaux2, Florian Tilquin2
1Laboratoire Traitement du Signal et de l'Image (LTSI-Inserm UMR 1099), Université de Rennes 1, 35042 Rennes, France.
Wearable devices like the Bora Band offer reliable heart rate variability (HRV) measurements. Ultra-short-term HRV analysis using photoplethysmography (PPG) is a valid surrogate for short-term analysis, aiding disease prevention.
Area of Science:
- Biomedical Engineering
- Physiological Monitoring
- Wearable Technology
Background:
- Continuous heart rate variability (HRV) monitoring via wearables aids physiological status assessment and disease prevention.
- Evaluating the accuracy of photoplethysmography (PPG) devices against electrocardiography (ECG) is crucial for reliable HRV analysis.
Purpose of the Study:
- To assess the agreement of HRV features between a commercial PPG wearable (Bora Band) and reference ECG.
- To validate ultra-short-term HRV measurements as a surrogate for short-term HRV analysis.
Main Methods:
- Acquired simultaneous PPG (Bora Band) and ECG recordings from 5 healthy subjects over 18 nights.
- Extracted time-domain, frequency-domain, nonlinear, and visibility graph HRV features from 5 min 30 s and 1 min 30 s durations.
- Compared PPG-derived HRV features against ECG reference using correlation, Bland-Altman plots, Cliff's delta, and equivalence testing.
Main Results:
- 37 out of 48 HRV features showed agreement between PPG and ECG in short-term recordings.
- 16 HRV features demonstrated strong correlation, minimal bias, and statistical equivalence in ultra-short (1 min 30 s) recordings.
- The Bora Band demonstrated valid and reliable HRV measurements across short and ultra-short durations.
Conclusions:
- The Bora Band provides accurate and dependable HRV measurements using PPG technology.
- Ultra-short-term HRV analysis is a valid approach, comparable to short-term analysis, for physiological monitoring.
- Wearable PPG devices show promise for accessible, continuous health status monitoring and disease prevention.