Related Experiment Video
Updated: Aug 23, 2025

Endotracheal Intubation Using a Flexible Intubation Endoscope as a Standardized Model for Safe Airway Management in Swine
Published on: August 25, 2022
A Comparative Evaluation of 2 Videolaryngoscopes as an Intubation Aid in a Simulated Difficult Airway: A Prospective
Ajay Kumar1, Anupma Taluja1, Bhavna Saxena1
1Department of Anaesthesia and Critical Care, Deen Dayal Upadhyay Hospital, New Delhi, India.
Objective:
Several videolaryngoscopes have been developed for using in difficult airway. We conducted this study to evaluate the performance of McGrath® MAC and King Vision® videolaryngoscopes in a simulated difficult airway.
Methods:
This prospective, randomised, comparative study was conducted in 140 surgical patients. Anaesthesia was administered as per standard protocol. A cervical collar was applied to simulate a difficult airway. Patients were randomised into 2 groups. In group M (n=70), laryngoscopy was performed first with King Vision® videolaryngoscope and second time with McGrath® MAC videolaryngoscope and trachea was intubated using the second device, while in group K (n=70), laryngoscopy was performed first with McGrath® MAC videolaryngoscope and second time with King Vision® videolaryngoscope and trachea was intubated using the second device. The laryngeal view, time taken for optimal laryngeal view, number of intubation attempts, ease of intubation, first attempt intubation success, time to tracheal intubation, haemodynamic parameters, and complications such as airway trauma, if any, were noted.
Results:
Tracheal intubation was faster with McGrath® MAC (34.89 ± 3.7 seconds) compared to King Vision® videolaryngoscope (43.43 ± 4.3 seconds, P <.001) with comparable first attempt intubation success by 100% vs 97.1%, P =.496, respectively. The laryngeal view obtained with both the devices was comparable but the mean time taken for optimal laryngeal view was significantly longer with King Vision® videolaryngoscope, both in group M (P <.001) and group K (P <.001). Ease of intubation and complications were comparable in the 2 groups.
Conclusion:
McGrath® MAC videolaryngoscope in comparison to King Vision® videolaryngoscope resulted in a shorter time taken for optimal laryngeal view and time to tracheal intubation with comparable first attempt intubation success.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Tracheostomy Suctioning II: Procedure