Rethinking Pulse Oximetry Screening in the Level-IV Neonatal Intensive Care Unit
Bryanna N Schwartz1,2, Lisa A Hom1,2, Mary E Revenis2,3
1Division of Cardiology, Children's National Heart Institute, Children's National Hospital, Washington, District of Columbia.
Objective:
This study aimed to evaluate the outcomes of newborn pulse oximetry screening in a level IV, tertiary care neonatal intensive care unit (NICU).
Study Design:
This is a retrospective cohort study of neonates who received newborn pulse oximetry screening after being admitted to a single-center, level-IV NICU between 2014 and 2021. Neonates with known critical congenital heart disease were excluded from the study.
Results:
Of the 4,493 neonates who had pulse oximetry screening, there were three positive screens (fail rate of 0.067%, 0.67 per 1,000 screened). The average age of screening was 818 hours. There were no positive screens of newborns who were admitted during their initial birth hospitalization and were screened while off oxygen. There were no new diagnoses of critical congenital heart disease (true positives) and there were no known false negatives.
Conclusion:
The results bring into question whether pulse oximetry screening with the current AAP-endorsed algorithm should be re-evaluated for a level-IV NICU at a children's hospital. However, the results may not be generalizable to other NICU's where echocardiography and prenatal echocardiograms are not as readily available.
Key Points:
· Pulse oximetry has been shown to be effective in decreasing delayed diagnosis of critical congenital heart disease (CCHD); however, there are limited prior studies on newborn pulse oximetry in the NICU.. · In our study of over 4,000 neonates admitted to a level IV tertiary care NICU, there were no true positives (no new diagnoses of CCHD).. · Special considerations may be needed for pulse oximetry screening in the NICU setting..
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
04:10Author Spotlight: Exploring Venous Waveforms for Non-Invasive Respiratory Monitoring in Pigs
Published on: March 8, 2024
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...