Evaluation of Analytical Performances and Comparison of 3 NT-proBNP Assays for Diagnosing Heart Failure
Jooyoung Cho1,2, Jong-Han Lee1, Sang-Guk Lee2
1From the Department of Laboratory Medicine, Yonsei University Wonju College of Medicine, Wonju, Korea (Cho, J.-H. Lee).
Context.—:
The N-terminal prohormone of the brain natriuretic peptide (NT-proBNP) is a major diagnostic biomarker for heart failure.
Objective.—:
To compare the analytical and clinical performance of 3 NT-proBNP immunoassays: the Atellica IM NT-proBNP assay (Siemens Healthcare Diagnostics), the Alere NT-proBNP assay (Abbott Laboratories), and the Elecsys proBNP II assay (Roche Diagnostics).
Design.—:
For the Atellica IM NT-proBNP assay, analytical performance, including precision, linearity, and carryover, was fully evaluated. Method comparisons among the 3 assays were performed using the Passing-Bablok regression and the κ agreement test. To evaluate the clinical performance of the assays, 160 patient samples were used from patients with (n = 81) or without (n = 79) heart failure.
Results.—:
The analytical performance of the Atellica IM NT-proBNP assay was acceptable according to the manufacturer's claims. The Atellica IM NT-proBNP assay showed a positive bias compared with the Elecsys proBNP II assay. The Cohen κ values among the 3 assays were satisfactory (>0.80) and comparable. There were no significant differences in areas under the curve. However, for the diagnosis of heart failure, the Elecsys proBNP II showed a higher specificity and positive likelihood ratio than the other assays.
Conclusions.—:
All 3 NT-proBNP assays showed acceptable concordance, and their clinical performance was comparable. However, the Elecsys proBNP II might be a more discriminating NT-proBNP assay to diagnose heart failure.
More Related Videos
08:21Functional Assessment of the Donor Heart During Ex Situ Perfusion: Insights from Pressure-Volume Loops and Surface Echocardiography
Published on: October 11, 2022
10:12A Sensitive and Specific Quantitation Method for Determination of Serum Cardiac Myosin Binding Protein-C by Electrochemiluminescence Immunoassay
Published on: August 8, 2013
Related Concept Videos
Heart Failure IV: Classification and Diagnostic Evaluation
Blood Studies for Cardiovascular System II: CRP, Hcy, and Cardiac Natriuretic Peptide Markers
These markers indicate stress or strain on the heart muscle:
Natriuretic Peptides (BNP)
Cardiac myocytes produce these hormones in response to ventricular stretching...
Acute Coronary Syndrome III: Diagnostic Studies
Dysrhythmias V: Evaluating Dysrhythmias
Heart Failure II: Pathophysiology
Heart Failure VII: Nursing Interventions