Sample Preparation for Probe Electrospray Ionization Mass Spectrometry
Published on: February 19, 2020
Coated Blade Spray with a Barrier: Improving Negative Electrospray Ionization and Sample Preparation
1Department of Chemistry, University of Waterloo, Waterloo, OntarioN2L 3G1, Canada.
A novel coated blade spray (CBS) design with a barrier significantly improves electrospray ionization stability for analyzing drugs in complex biological samples. This enhanced method achieves high reproducibility and accuracy in negative and positive ionization modes without internal standards.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Forensic Science
Background:
- Traditional substrate-based electrospray ionization (ESI) methods struggle with unstable spray, high noise, and corona discharge in negative ionization mode, particularly for complex biological matrices like blood and urine.
- Coated blade spray coupled with mass spectrometry (CBS-MS) offers an alternative, combining solid-phase microextraction (SPME) efficiency with ambient MS speed for rapid analyte screening.
Purpose of the Study:
- To develop and validate a new coated blade spray (CBS) device with a barrier feature to enhance electrospray ionization (ESI) stability and performance in negative ionization mode.
- To assess the reproducibility and accuracy of the new CBS design for the analysis of drugs of abuse in complex biological samples, specifically urine and plasma.
Main Methods:
- Development of a modified CBS blade incorporating a barrier at the ESI tip to mitigate instability issues.
- Investigation of ESI stability using microscopic observation and CBS-MS testing.
- Application of the new CBS device for the quantitative analysis of drugs of abuse in urine (negative mode) and plasma (positive mode) without internal standards.
Main Results:
- The addition of a barrier to the CBS blade reduced the total relative standard deviation (RSD%) to below 10% for sample preparation, ionization, and MS detection of drugs in negative mode.
- Analysis of drugs in urine yielded high calibration curve correlation coefficients (R² ≥ 0.9997) and an overall peak area RSD% of 6.9% for one compound across 42 samples.
- On-blade sampling of 14 drugs in plasma (positive mode) showed R² ≥ 0.9883 and limits of quantification (LOQs) ranging from 0.25 to 25 ng/mL.
Conclusions:
- The novel CBS blade design with a barrier significantly enhances ESI stability and reproducibility, particularly in negative ionization mode, for analyzing drugs in complex biological matrices.
- The method demonstrates excellent quantitative performance without the need for internal standards, highlighting its potential for high-throughput screening and automation.
- The findings offer valuable insights for optimizing CBS device design and automating protocols for ambient mass spectrometry applications.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Preparation of Samples for Electron Microscopy
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Capillary Electrophoresis: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...