Primary Measurement of Gaseous Elemental Mercury Concentration with a Dynamic Range of Six Decades
Abneesh Srivastava1, Joseph T Hodges1
1Chemical Sciences Division, Material Measurement Laboratory, National Institute of Standards and Technology (NIST), 100 Bureau Drive, Gaithersburg, Maryland20899, United States.
Abstract:
We present a primary method for absolute reference material-free measurement of gaseous elemental mercury mass concentration based on laser absorption spectroscopy at wavelengths near 254 nm and anchored to the international system of units. A 76 m-long multipass cell provides a detection limit of 1 ng m-3 for an averaging time of 8 min. A dual-cell arrangement, comprising a single-pass and a multipass cell, provides a dynamic range of six decades and spans from ambient to emission levels of gaseous elemental mercury (1 ng m-3 to 1000 μg m-3). The relative standard uncertainty of the measurement is 0.5% in the 0.25 to 1000 μg m-3 range.
More Related Videos
Related Concept Videos
Voltammetric Techniques: Linear-Scan (E vs Time)
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer