Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Gas Chromatography–Mass Spectrometry (GC–MS)01:14

Gas Chromatography–Mass Spectrometry (GC–MS)

4.5K
Gas chromatography–mass spectrometry (GC–MS) is the combination of analytical techniques of gas chromatography and mass spectrometry in a single instrument for analyzing a mixture of compounds. The gas chromatograph separates the compounds in the mixture, and the mass spectrometer analyzes each compound separately to determine the molecular masses and molecular structures.
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
4.5K
Gas Chromatography: Sample Injection Systems01:08

Gas Chromatography: Sample Injection Systems

501
In gas chromatography, the sample is introduced as a vapor plug into the carrier gas stream for high efficiency and resolution. A microsyringe injects the sample solution into a heated sample port, vaporizing it and mixing it with the carrier gas. This process is important to ensure the sample is properly prepared for analysis. Thermally sensitive samples can be injected directly into the column and volatilized by slowly increasing the column temperature.
Two primary injection methods are used...
501
Gas Chromatography: Types of Detectors-II01:19

Gas Chromatography: Types of Detectors-II

477
In gas chromatography, different detectors are employed to meet specific analytical needs. These detectors are often categorized based on their detection mechanisms and the types of compounds they are best suited to analyze. Thermal Conductivity Detectors (TCD), Flame Ionization Detectors (FID), and Electron Capture Detectors (ECD) represent common categories, each with unique operating principles and applications. However, beyond these, several other detectors are designed for more specialized...
477
Gas Chromatography: Introduction01:13

Gas Chromatography: Introduction

2.2K
Gas chromatography (GC) is a technique for separating and analyzing volatile compounds in a sample. Its primary purpose is to identify and quantify components in complex mixtures, making it essential in fields such as environmental analysis, pharmaceuticals, and petrochemicals. GC is also called vapor-phase chromatography (VPC) or gas-liquid partition chromatography (GLPC).
In GC,  a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
2.2K
Gas Chromatography: Overview of Detectors01:13

Gas Chromatography: Overview of Detectors

701
Detectors in gas chromatography (GC) help identify and quantify the components of a mixture by translating chemical properties into measurable signals, which are displayed on a chromatogram. Detectors can be categorized into two main types: destructive and non-destructive.
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
701
Gas Chromatography: Types of Detectors-I01:21

Gas Chromatography: Types of Detectors-I

561
There are different types of detectors used in gas chromatography, each with its own specific properties that make it suitable for detecting certain types of analytes. The most commonly used detectors in GC are thermal conductivity detector (TCD), flame ionization detector (FID), and electron capture detector (ECD).
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
561

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Integrated Transcriptomic and Metabolomic Analyses Reveal the Mechanism by Which 5-Methoxyindole Enhances Sesquiterpenoids Production in <i>Atractylodes chinensis</i> Hairy Roots.

Plants (Basel, Switzerland)·2026
Same author

A peripheral subpopulation of retinal pigment epithelium resists oxidative damage through SERPINE3-mediated Caspase-1 inhibition.

The Journal of clinical investigation·2026
Same author

Protocol: Comparative efficacy of 2 mL vs. 5 mL 0.375% ropivacaine in subparaneural upper trunk brachial plexus block for postoperative analgesia after arthroscopic shoulder surgery-a randomized noninferiority trial.

Frontiers in medicine·2026
Same author

Resistant hypertension: Epidemiology, mechanisms and contemporary therapeutic strategies.

Heart international·2026
Same author

Serum uric acid in systemic lupus erythematosus with preserved renal function: a cross-sectional and longitudinal analysis.

Frontiers in immunology·2026
Same author

Gene therapy induces pigmentation in induced pluripotent stem cell-derived retinal pigment epithelium of OCA1A patients and albino rat eye.

Molecular therapy. Advances·2026

Related Experiment Video

Updated: Aug 23, 2025

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
07:57

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector

Published on: July 25, 2014

20.1K

Portable comprehensive two-dimensional micro-gas chromatography using an integrated flow-restricted pneumatic

Xiaheng Huang1,2,3,4, Maxwell Wei-Hao Li1,2,3,4, Wenzhe Zang1,3

  • 1Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI 48109 USA.

Microsystems & Nanoengineering
|November 4, 2022
PubMed
Summary

A novel flow-restricted pneumatic modulator (FRPM) enables rapid 2D injection for portable comprehensive 2D micro-GC (μGC). This technology achieves fast separation of complex volatile organic compounds, enhancing analytical capabilities.

Keywords:
ChemistryElectrical and electronic engineering

More Related Videos

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
11:44

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry

Published on: March 6, 2016

9.4K
Design and Use of a Full Flow Sampling System FFS for the Quantification of Methane Emissions
08:18

Design and Use of a Full Flow Sampling System FFS for the Quantification of Methane Emissions

Published on: June 12, 2016

16.9K

Related Experiment Videos

Last Updated: Aug 23, 2025

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
07:57

Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector

Published on: July 25, 2014

20.1K
Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry
11:44

Qualitative Characterization of the Aqueous Fraction from Hydrothermal Liquefaction of Algae Using 2D Gas Chromatography with Time-of-flight Mass Spectrometry

Published on: March 6, 2016

9.4K
Design and Use of a Full Flow Sampling System FFS for the Quantification of Methane Emissions
08:18

Design and Use of a Full Flow Sampling System FFS for the Quantification of Methane Emissions

Published on: June 12, 2016

16.9K

Area of Science:

  • Analytical Chemistry
  • Chromatography
  • Microfluidics

Background:

  • Conventional one-dimensional gas chromatography (GC) faces limitations in separating highly complex chemical samples.
  • Two-dimensional (2D) GC offers enhanced separation but often requires complex instrumentation.
  • Portable analytical devices are increasingly important for on-site chemical analysis.

Purpose of the Study:

  • To develop a microfabricated flow-restricted pneumatic modulator (FRPM) for portable comprehensive 2D micro-GC (μGC).
  • To enable rapid 2D injection and separation without compromising 1D separation speed and peak profiles.
  • To demonstrate the performance and application of the integrated μGC system for complex sample analysis.

Main Methods:

  • Development and characterization of a microfabricated FRPM for 2D injection.
  • Integration of the FRPM with 1D and 2D micro-columns on a single chip.
  • Utilized flow-through micro-photoionization detectors (μPIDs) for injection characteristic analysis.
  • Assembled an automated portable comprehensive 2D μGC system.

Main Results:

  • Achieved a 2D injection peak width of approximately 25 ms with a 2D/1D flow rate ratio over 10.
  • Successfully integrated a 0.5-m 2D μcolumn with the FRPM on-chip.
  • Demonstrated rapid separation of 40 volatile organic compounds in approximately 5 minutes using the portable 2D μGC.
  • Constructed a hybrid 2D contour plot using data from two μPIDs.

Conclusions:

  • The developed FRPM is effective for rapid 2D injection in portable μGC systems.
  • The integrated 2D μGC system provides fast and efficient separation of complex volatile organic compound mixtures.
  • The FRPM design facilitates enhanced analytical performance for comprehensive 2D GC applications.