Related Experiment Video
Updated: Aug 23, 2025

Methylation Specific Multiplex Droplet PCR using Polymer Droplet Generator Device for Hematological Diagnostics
Published on: June 29, 2020
Development and validation of methods that enable high-quality droplet digital PCR and hematological profiling data
Bin Xu1, Ryan Haley1
1Biologics Team, Frontage Laboratories, 760 Pennsylvania Drive, Exton, PA 19341, USA.
Abstract:
Aim: Mouse models have been crucial to preclinical studies in the increasingly relevant fields of cell and gene therapy. However, only small quantities of mouse blood can be collected without producing adverse physiological effects that compromise data integrity. Results: To address this limitation, two combined methods were developed to create detailed droplet digital PCR (ddPCR) and hematological profiles using only ∼20 μl of mouse blood. The validation of these methods, which can serve as a foundation for a standardized regulatory pipeline for ddPCR, is discussed. Even when using small amounts of input, this ddPCR protocol is accurate, precise, selective, specific, stable and robust. Conclusion: These techniques enable more frequent sample collection for higher-resolution pharmacokinetic data that meets or exceeds quality standards.