Contrast thresholds for detection of various iodine concentrations in subtraction CT and dual-energy CT systems
Anahita Heshmat1, Izabella Barreto1, Lynn Rill1
1Department of Radiology, College of Medicine, University of Florida, Gainesville, Florida, USA.
Objective:
To estimate the minimum iodine concentrations detectable in simulated vessels of various diameters for both subtraction computed tomography (CT) and dual-energy CT systems.
Methods:
Fillable tubes (diameters: 1, 3, and 5 mm) were filled with a variety of iodine concentrations (range: 0-20 mg/ml), placed in the center of 28-mm cylindrical rods and surrounded with water. Rods with and without fillable tubes were placed in a 20-cm cylindrical solid-water phantom to simulate administration of iodine in blood vessels. The phantom was scanned with clinical subtraction CT (SCT) and dual-energy CT (DECT) head protocols to assess the detection of minimum iodine concentrations in both systems. The SCT and DECT images were evaluated quantitatively with a MATLAB script to extract regions of interest (ROIs) of each simulated vessel. ROI measurements were used to calculate the limit of detectability (LOD) and signal-to-noise ratio of Rose criteria for the assessment of the contrast thresholds.
Results:
Both SNRRose and LOD methods agreed and determined the minimum detectable iodine concentration to be 0.4 mg/ml in the 5-mm diameter vessel for SCT. However, the minimum detectable concentration in the 5-mm vessel with DECT was 1 mg/ml. The 3-mm vessel had a minimum detectable concentration of 0.8 mg/ml for SCT and 2 mg/ml for DECT. Lastly, the minimum detectable iodine concentration for the 1-mm vessel was 10 mg/ml for SCT and 10 mg/ml for DECT.
Conclusion:
In this phantom study, SCT showed the capability to detect lower iodine concentrations compared to DECT. Contrast thresholds varied for vessels of different diameters and the smaller vessels required a higher iodine concentration for detection. Based on this knowledge, radiologists can modify their protocols to increase contrast enhancement.
More Related Videos
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Redox Titration: Iodimetry and Iodometry
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...