Related Experiment Video
Updated: Aug 23, 2025

Three-Dimensional Preoperative Virtual Planning in Derotational Proximal Femoral Osteotomy
Published on: February 17, 2023
3D-planned corrective osteotomy of metacarpal arch disruption after Motec® wrist prosthesis
A Durdzińska Timóteo1, T Voser1, S Durand1
1Centre Hospitalier Universitaire Vaudois, Department of Plastic and Hand Surgery, Avenue Pierre-Decker 4, 1005 Lausanne, Switzerland.
Abstract:
We report two cases of metacarpal arch disruption after Motec® total wrist arthroplasty, with hyperextension of the third metacarpal and dorsal protrusion of the head. Correction osteotomies of the adjacent metacarpals using preoperative 3D surgical planning were successful. This report seeks to shed light on the origin of this deformation and to recommend some operative precautions. It is important that surgeons should be aware of the existence of this complication, as disruption of the transverse metacarpal arch affects both grasp and hand esthetics. LEVEL OF EVIDENCE: IV.