Comprehensive clinical evaluation of TomoEQA for patient-specific pre-treatment quality assurance in helical
Changhwan Kim1, Min Cheol Han2, Young Kyu Lee3
1Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul, 03722, Republic of Korea.
Background:
Based on a previous study on the feasibility of TomoEQA, an exit detector-based patient-specific pre-treatment quality assurance (QA) method for helical tomotherapy, an in-depth clinical evaluation was conducted.
Methods:
Data of one hundred patients were analyzed to evaluate the clinical usefulness of TomoEQA for patient-specific pre-treatment QA in comparison with the conventional phantom-based method. Additional investigations were also performed under unusual measurement conditions to validate the off-axis region. In addition to the clinical evaluation of TomoEQA, a statistical analysis was conducted to determine the plan parameters that affect the pass/failure results of pre-treatment QA.
Results:
The average and standard deviations of the gamma passing rate and point dose error for TomoEQA were comparable to those of the conventional QA method. For TomoEQA, the average values of the gamma passing rate and point dose error were 96.32% (standard deviation (1 sigma) = 3.94; 95% confidence interval (CI), 95.55 to 97.09) and - 1.12% (standard deviation (1 sigma) = 1.04; CI, - 1.32 to - 0.92), respectively. For the conventional QA method, the average values of the gamma passing rate and point dose error were 95.95% (standard deviation (1 sigma) = 4.35; 95% confidence interval (CI), 95.10 to 96.80) and - 1.20% (standard deviation (1 sigma) = 1.61; CI, - 1.52 to - 0.88), respectively. Further experiments on the off-axis region demonstrated that TomoEQA can provide accurate results for 3D dose analysis, which is inherently difficult in the conventional QA method. Through a statistical analysis based on the results of TomoEQA, it was validated that the total fraction (Total Fx), monitor units, beam-on-time, leaf-of-time below 100 ms, and planning target volume diameter were statistically significant for the pass/failure of the pre-treatment QA results.
Conclusions:
TomoEQA is a clinically beneficial alternative to the conventional phantom-based QA method.