Test-retest reliability of splenic volume assessment by ultrasonography
Pontus Holmström1, Frank Pernett2, Erika Schagatay2,3
1Environmental Physiology Group, Department of Health Sciences, Mid Sweden University, Kunskapsgatan 4, 83140, Östersund, Sweden. Pontus.holmstrom@miun.se.
Abstract:
While MRI and CT are the gold standards for assessments of splenic size in clinical settings, ultrasonography is particularly suited due to its portability, cost efficiency and easy utilization. However, ultrasonography is associated with subjective assessment, potentially resulting in increased variation. We used a test-retest design aiming to determine the reliability of splenic measurements assessed by ultrasonography during apnea. In addition, we compared reliability between different equations for volume calculations: Koga, Prolate ellipsoid and Pilström. Twelve healthy participants (6 women) performed two tests separated by 15 min, comprising a maximal voluntary apnea in a seated position. Splenic dimensions were measured via ultrasonography for 5 min before and immediately following apnea. Resting splenic volume displayed high test-retest reliability between tests (Pilström: 157 ± 39 mL vs 156 ± 34 mL, p = .651, ICC = .970, p < .001, CV = 2.98 ± 0.1%; Prolate ellipsoid: 154 ± 37 mL vs 144 ± 43 mL, p = .122, ICC = .942, p < .001, CV = 5.47 ± 0.3%; Koga: 142 ± 37 mL vs 140 ± 59 mL, p = .845, ICC = .859, p < .001, CV = 9.72 ± 1.4%). Apnea-induced volumes displayed similar reliability (127 ± 29 mL vs 129 ± 28 mL, p = .359, ICC = .967, p < .001, CV = 3.14 ± 3.1%). Reliability was also high between equations (Pilström vs Prolate ellipsoid: ICC = .818, p < .001, CV = 7.33 ± 0.3%, bias = - 3.1 mL, LoA = - 46.9 to 40.7 mL; Pilström vs Koga: ICC = .618, p < .01, CV = 11.83 ± 1.1%, bias = - 14.8 mL, LoA = - 76.9 to 47.3 mL). We conclude that splenic ultrasonographic measurements have practical applications during laboratory and field-based research as a reliable method detecting splenic volume change consistently between repeated tests. The Pilström equation displayed similar reliability compared to the prolate ellipsoid formula and slightly higher compared to the Koga formula and may be particularly useful to account for individual differences in splenic dimensions.
More Related Videos
07:03Imaging and Quantification of the Hepatic Vasculature of Mice Using Ultrafast Doppler Ultrasound
Published on: July 19, 2024
10:38Observational Study Protocol for Repeated Clinical Examination and Critical Care Ultrasonography Within the Simple Intensive Care Studies
Published on: January 16, 2019
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...