Tejas Kalaria1, Clare Ford1, Rousseau Gama1,2
1Clinical Biochemistry, New Cross Hospital, 592016Black Country Pathology Services, The Royal Wolverhampton NHS Trust, Wolverhampton, UK.
Background:
The widely accepted practice of not reporting any results from ethylenediaminetetraacetic acid (EDTA) contaminated samples necessitates repeat phlebotomy and could delay clinical management decisions. EDTA, however, interferes variably or not at all in assays. EDTA concentration in contaminated samples, like serum indices, therefore, could be used to selectively report the result of analytes not affected at measured EDTA concentration.
Methods:
A serum pool, level 1 and level 3 internal quality control materials were spiked with tripotassium-EDTA to create samples with EDTA concentration up to 6.0 mmol/L. EDTA and 45 common and critically important analytes were measured on Abbott Architect to identify EDTA concentrations for analytes where the change in concentration exceeded their respective reference change value (RCV) for unidirectional change at 95% probability.
Results:
Serum potassium increased and calcium decreased exceeding RCV at 0.17 mmol/L EDTA. Alkaline phosphatase (ALP) decreased exceeding RCV at EDTA >1.86 mmol/L. The decrease in iron did not exceed a wide RCV of 61.9% until maximum spiked EDTA but exceeded the desirable specification for allowable total error (30.7%) at EDTA >1.86 mmol/L. The small decrease in magnesium did not exceed RCV. EDTA up to the concentration in blood collection tubes did not affect the results of any other measured analyte.
Conclusions:
Only serum potassium, calcium, ALP and iron studies, of the 45 analytes studied, should not be reported in EDTA contaminated samples. EDTA concentration cut-offs for selective reporting would further facilitate reporting of these analytes in EDTA contaminated samples.
More Related Videos
07:54Analysis of AtHIRD11 Intrinsic Disorder and Binding Towards Metal Ions by Capillary Gel Electrophoresis and Affinity Capillary Electrophoresis
Published on: August 22, 2018
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
Related Concept Videos
EDTA: Auxiliary Complexing Reagents
Masking and Demasking Agents
There are many masking agents, such as cyanide, fluoride, triethanolamine, thiourea, and 2,3-bis(sulfanyl)propan-1-ol (formerly 2,3-dimercapto-1-propanol), with the masking agent chosen based on...
EDTA: Chemistry and Properties
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...
EDTA: Indirect and Alkalimetric Titration