Canceling Notch Improves the Mechanical Safety of Clavicle Locking Plate: A 3D Finite Element Study
Zhihua Han1,2, Qian Luo3, Guoying Deng1
1Trauma Center, Department of Orthopaedics and Traumatology, Shanghai General Hospital, Shanghai Jiaotong University, Shanghai, China.
Objective:
Implant failure is a disastrous complication of the operative treatment of midshaft clavicle fractures, and improving the osteosynthesis plate is a strategy for preventing this. We aimed to investigate whether canceling the notch and adding screw-hole inserts enhanced the mechanical properties of the plate.
Methods:
A clavicle model was generated based on the CT images of six adult volunteers (age range, 20-40 years; three males and three females; height range 160-175) using dedicated software, and a midshaft fracture model was created. The domestically made seven-hole locking plate commonly used for midshaft clavicle fractures was simulated (Model I); modifications were made to the plate (Model II). Using 3D finite element analysis, we simulated the fracture construct under three different load conditions-downward cantilever bending, axial compression, and axial torsion-and compared the stress distribution.
Results:
We found that under axial compression, Model II experienced its maximum stress on the plate at 551.9MPa, which was less than that in Model I (790.4 MPa). Moreover, a greater stress concentration at the fracture site was observed under axial torsion, despite the maximum stress of both the models being similar.
Conclusion:
Canceling the notch and filling the screw holes near the fracture can ameliorate stress concentration on the internal fixation construct and enhance its reliability under axial compression. This improvement has substantial effects on the mechanical properties of implants and potentially prevents implant failure. Modern osteosynthesis anatomical implants need to be improved.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
06:38Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
Related Concept Videos
Machines: Problem Solving I
The toggle clamp system is a machine structure consisting of movable, pin-connected multi-force members that form a stabilized system to transmit forces. The...
Changes in the Appendicular Skeleton with Age
Initially, the limb buds consist of a core of mesenchyme covered by a layer of ectoderm. The ectoderm at the end of the limb bud thickens to form a narrow crest called the apical ectodermal ridge. This ridge stimulates the underlying...