Myeloma bone disease imaging on a 1st-generation clinical photon-counting detector CT vs. 2nd-generation dual-source
Moritz T Winkelmann1, Florian Hagen2, Lucie Le-Yannou2
1Department for Diagnostic and Interventional Radiology, University Hospital Tübingen, Hoppe-Seyler-Straße 3, 72076, Tübingen, Germany. moritz.winkelmann@med.uni-tuebingen.de.
Objective:
Subjective and objective image quality comparison of bone microstructure and disease-related abnormalities in multiple myeloma patients using a 1st-generation dual-source photon-counting detector CT(DS-PCD-CT) and a 2nd-generation dual-source dual-energy (energy-integrating detector) CT (DS-EID-CT).
Methods:
Fifty multiple myeloma patients (mean age 67.7 ± 10.9 years,16 females) were prospectively enrolled. Unenhanced whole-body CTs were clinically indicated and performed on DS-EID-CT and DS-PCD-CT (median time difference: 12 months). DS-PCD-CT was performed in Quantumplus UHR mode and DS-EID-CT was performed using dual-energy mode. DS-PCD-CT kernel was set at Br64 with Quantum iterative reconstruction strength Q1; for DS-EID-CT a comparable I70f kernel with SAFIRE iterative reconstruction strength 1 was used. Two independent radiologists assessed image quality subjectively using a 5-point Likert scale considering delineation and sharpness of trabecular bone and lytic bone lesions in the spine and pelvic bones. Additionally, ImageJ was used for quantification of bony septa inside the cancellous bone and through or the edges of osteolysis.
Results:
Overall quality as well as detectability and sharpness in the delineation of lytic bone lesions were superior for DS-PCD-CT compared with DS-EID-CT (p < 0.0001). The inter-reader agreement for subjective image quality readings showed excellent consistency(α = 94.2-98.8). CTDI and DLP mean values for DS-PCD-CT and DS-EID-CT were 1107.4 ± 247.6 mGy*cm and 8.2 ± 1.8 mGy vs. 1344.3 ± 204.6 mGy*cm and 10.1 ± 1.9 mGy. The quantitative metric for bone microstructure in the femoral head showed significantly better visualization of trabeculae in DS-PCD-CT compared with DS-EID-CT (p < 0.0001). Quantitative analyses of edge sharpness of osteolysis showed significant steeper edges for DS-PCD-CT (p < 0.0001).
Conclusion:
DS-PCD-CT significantly improves spatial resolution of bony microstructure and lytic bone lesions compared to DS-EID-CT.
Key Points:
• Application of photon-counting detector CT is superior to dual-source dual-energy integrating detector in clinical workup of multiple myeloma patients. • Compared to energy integrating detectors, photon-counting detectors significantly increase the spatial resolution of bone microstructure including disease-related lytic bone lesions in patients with multiple myeloma.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET