CT image quality evaluation in the age of deep learning: trade-off between functionality and fidelity
Kai Yang1, Jinjin Cao2, Nisanard Pisuchpen2
1Division of Diagnostic Imaging Physics, Department of Radiology, Massachusetts General Hospital, 55 Fruit Street, Boston, MA, 02114, USA. kyang11@mgh.harvard.edu.
Objective:
To quantitatively compare DLIR and ASiR-V with realistic anatomical images.
Methods:
CT scans of an anthropomorphic phantom were acquired using three routine protocols (brain, chest, and abdomen) at four dose levels, with images reconstructed at five levels of ASiR-V and three levels of DLIR. Noise power spectrum (NPS) was estimated using a difference image by subtracting two matching images from repeated scans. Using the max-dose FBP reconstruction as the ground truth, the structure similarity index (SSIM) and gradient magnitude (GM) of difference images were evaluated. Image noise magnitude (σ), frequency location of the NPS peak (fpeak), mean SSIM (MSSIM), and mean GM (MGM) were used as quantitative metrics to compare image quality, for each anatomical region, protocol, algorithm, dose level, and slice thickness.
Results:
Image noise had a strong (R2 > 0.99) power law relationship with dose for all algorithms. For the abdomen and chest, fpeak shifted from 0.3 (FBP) down to 0.15 mm-1 (ASiR-V 100%) with increasing ASiR-V strength but remained 0.3 mm-1 for all DLIR levels. fpeak shifted down for the brain protocol with increasing DLIR levels. Three levels of DLIR produced similar image noise levels as ASiR-V 40%, 80%, and 100%, respectively. DLIR had lower MSSIM but higher MGM than ASiR-V while matching imaging noise.
Conclusion:
Compared to ASiR-V, DLIR presents trade-offs between functionality and fidelity: it has a noise texture closer to FBP and more edge enhancement, but reduced structure similarity. These trade-offs and unique protocol-dependent behaviors of DLIR should be considered during clinical implementation and deployment.
Key Points:
• DLIR reconstructed images demonstrate closer noise texture and lower structure similarity to FBP while producing equivalent noise levels comparable to ASiR-V. • DLIR has additional edge enhancement as compared to ASiR-V. • DLIR has unique protocol-dependent behaviors that should be considered for clinical implementation.
Related Concept Videos
Improving Translational Accuracy
Imaging Studies III: Computed Tomography
Three-Dimensional Microscopy in Microbiology
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...