Development and Optimization of Silicon-Dioxide-Coated Capacitive Electrode for Ambulatory ECG Measurement System
Younghwan Kang1, Sangdong Choi1, Chiwan Koo1
1Department of Electronic Engineering, Hanbat National University, Yuseong-gu, Daejeon 34158, Korea.
Abstract:
This paper presents a silicon-dioxide-coated capacitive electrode system for an ambulatory electrocardiogram (ECG). The electrode was coated with a nano-leveled (287 nm) silicon dioxide layer which has a very high resistance of over 200 MΩ. Due to this high resistance, the electrode can be defined as only a capacitor without a resistive characteristic. This distinct capacitive characteristic of the electrode brings a simplified circuit analysis to achieve the development of a high-quality ambulatory ECG system. The 240 um thickness electrode was composed of a stainless-steel sheet layer for sensing, a polyimide electrical insulation layer, and a copper sheet connected with the ground to block any electrical noises generated from the back side of the structure. Six different diameter electrodes were prepared to optimize ECG signals in ambulatory environment, such as the amplitude of the QRS complex, amplitude of electromagnetic interference (EMI), and baseline wandering of the ECG signals. By combining the experimental results, optimal ambulatory ECG signals were obtained with electrodes that have a diameter from 1 to 3 cm. Moreover, we achieved high-quality ECG signals in a sweating simulation environment with 2 cm electrodes.
More Related Videos
05:57Author Spotlight: Microfluidic Channel-Based Soft Electrodes and Their Application in Capacitive Pressure Sensing
Published on: March 17, 2023
10:28Sensitivity Enhancement of Soft Capacitive Pressure Sensors Using a Solvent Evaporation-Based Porosity Control Technique
Published on: March 24, 2023