Analysis and Correction of Polarization Response Calibration Error of Limb Atmosphere Ultraviolet Hyperspectral
Haochen Li1,2, Zhanfeng Li1, Yu Huang1
1Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China.
Abstract:
A UV hyperspectral instrument was designed with a polarization measurement channel for real-time in-orbit polarization correction to reduce the influence of polarization on the detection accuracy of atmospheric radiation. One of the prerequisites for in-orbit polarization calibration is accurately calibrating the instrument's polarization properties in the laboratory. This study first introduces the calibration method and measuring device of the polarization characteristics of the ultraviolet (UV) hyperspectral detector and conducts a polarization calibration test of the instrument. The two main error sources introduced by the calibration device were emphatically analyzed, and the correction method of the error sources was deduced theoretically. Finally, the polarization calibration results of the UV hyperspectral detector were corrected, and the uncertainty analysis of the corrected calibration results was about 1.4%, which provides effective ground polarization calibration data for the on-orbit polarization correction of the instrument.
More Related Videos
10:22Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
Published on: September 7, 2019
09:00Indoor Experimental Assessment of the Efficiency and Irradiance Spot of the Achromatic Doublet on Glass ADG Fresnel Lens for Concentrating Photovoltaics
Published on: October 27, 2017
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
UV–Vis Spectrometers
UV–Vis Spectroscopy of Conjugated Systems
One of the factors influencing λmax is the extent...
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Flame Photometry: Lab
UV–Vis Spectroscopy: Beer–Lambert Law