The use of cardiac CT acquisition mode for dynamic musculoskeletal imaging
Benyameen Keelson1, Luca Buzzatti2, Gert Van Gompel3
1Vrije Universiteit Brussel (VUB), Universitair Ziekenhuis Brussels (UZB), Department of radiology, Laarbeeklaan 101, 1090 Brussels, Belgium; Vrije Universiteit Brussel (VUB), Department of Electronics and Informatics (ETRO), Pleinlaan 9, 1050 Brussels, Belgium; IMEC, Kapeldreef 75, B-3002 Leuven, Belgium.
Objectives:
To quantitatively evaluate the impact of a cardiac acquisition CT mode on motion artifacts in comparison to a conventional cine mode for dynamic musculoskeletal (MSK) imaging.
Methods:
A rotating PMMA phantom with air-filled holes drilled at varying distances from the disk center corresponding to linear hole speeds of 0.75 cm/s, 2.0 cm/s, and 3.6 cm/s was designed. Dynamic scans were obtained in cardiac and cine modes while the phantom was rotating at 48°/s in the CT scanner. An automated workflow to compute the Jaccard distance (JD) was established to quantify degree of motion artifacts in the reconstructed phantom images. JD values between the cardiac and cine scan modes were compared using a paired sample t-test. In addition, three healthy volunteers were scanned with both modes during a cyclic flexion-extension motion of the knee and analysed using the proposed metric.
Results:
For all hole sizes and speeds, the cardiac scan mode had significantly lower (p-value <0.001) JD values. (0.39 [0.32-0.46]) i.e less motion artifacts in comparison to the cine mode (0.72 [0.68-0.76]). For both modes, a progressive increase in JD was also observed as the linear speed of the holes increased from 0.75 cm/s to 3.6 cm/s. The dynamic images of the three healthy volunteers showed less artifacts when scanned in cardiac mode compared to cine mode, and this was quantitatively confirmed by the JD values.
Conclusions:
A cardiac scan mode could be used to study dynamic musculoskeletal phenomena especially of fast-moving joints since it significantly minimized motion artifacts.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT