Bedside Ultrasound for Guiding Fluid Removal in Patients with Pulmonary Edema: The Reverse-FALLS Protocol
Published on: July 28, 2018
MONITORING OF PULMONARY EDEMA USING ULTRASOUND RADIOFREQUENCY SIGNAL
Hyuksool Kwon, You Hwan Jo, Jae Hyuk Lee
1Department of Emergency Medicine, Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
Abstract:
Objectives: Excessive accumulation of extravascular lung water impairs respiratory gas exchange and results in respiratory distress. Real-time radiofrequency signals of ultrasound can continuously and quantitatively monitor excessive lung water. This study aims to evaluate the availability of continuous real-time quantitative pulmonary edema monitoring using ultrasound radiofrequency signals and compare it with Pa o2 (partial pressure of arterial oxygen)/F io2 (fraction of inspired oxygen) (PF) ratio, conventional lung ultrasound, and the Hounsfield unit of chest computed tomography. Methods: Male Yorkshire pigs (40.5 ± 0.5 kg) were anesthetized and mechanically ventilated. A balanced crystalloid was administered to induce hydrostatic pulmonary edema. Three different infusion rates of 2, 4, and 6 mL/kg per minute were tested to determine the infusion rate for the appropriate swine model. The chest computed tomography and ultrasonography with radiofrequency signals were taken every 5 min during the full inspiration. The ultrasonography scans with radiofrequency signals were measured at the intercostal space where the line crossing the two armpits and the right anterior axillary line intersected. Results: The infusion rate of fluid for the pulmonary edema model was determined to be 6 mL/kg per minute, and a total of four pigs were tested at an injection rate of 6 mL/kg. The adjusted R2 values of regression analysis between the radiofrequency signal and computer tomography Hounsfield score were 0.990, 0.993, 0.988, and 0.993 (all P values <0.05). All radiofrequency signal changes preceded changes in PF ratio or lung ultrasound changes. The area under the receiver operating characteristic curve of the radiofrequency signal for predicting PF ratio <300 was 0.88 (95% confidence interval, 0.82-0.93). Conclusion: We evaluated ultrasound radiofrequency signals to assess pulmonary edema in a swine model that can worsen gradually and showed that quantitative ultrasound radiofrequency signal analysis could assess pulmonary edema and its progression before PF ratio or lung ultrasound changes.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasonography
During an ultrasonography procedure, a handheld device called...