Related Experiment Video
Updated: Aug 21, 2025

Combining Volumetric Capnography And Barometric Plethysmography To Measure The Lung Structure-function Relationship
Published on: January 8, 2019
Experimental validation of a portable tidal volume indicator for bag valve mask ventilation
Benjamin S Maxey1, Luke A White1, Giovanni F Solitro2
1Department of Molecular & Cellular Physiology, LSU Health Shreveport, 1501 Kings Highway, Shreveport, LA, 71103-3932, USA.
Introduction:
Short-term emergency ventilation is most typically accomplished through bag valve mask (BVM) techniques. BVMs like the AMBU® bag are cost-effective and highly portable but are also highly prone to user error, especially in high-stress emergent situations. Inaccurate and inappropriate ventilation has the potential to inflict great injury to patients through hyper- and hypoventilation. Here, we present the BVM Emergency Narration-Guided Instrument (BENGI) - a tidal volume feedback monitoring device that provides instantaneous visual and audio feedback on delivered tidal volumes, respiratory rates, and inspiratory/expiratory times. Providing feedback on the depth and regularity of respirations enables providers to deliver more consistent and accurate tidal volumes and rates. We describe the design, assembly, and validation of the BENGI as a practical tool to reduce manual ventilation-induced lung injury.
Methods:
The prototype BENGI was assembled with custom 3D-printed housing and commercially available electronic components. A mass flow sensor in the central channel of the device measures air flow, which is used to calculate tidal volume. Tidal volumes are displayed via an LED ring affixed to the top of the BENGI. Additional feedback is provided through a speaker in the device. Central processing is accomplished through an Arduino microcontroller. Validation of the BENGI was accomplished using benchtop simulation with a clinical ventilator, BVM, and manikin test lung. Known respiratory quantities were delivered by the ventilator which were then compared to measurements from the BENGI to validate the accuracy of flow measurements, tidal volume calculations, and audio cue triggers.
Results:
BENGI tidal volume measurements were found to lie within 4% of true delivered tidal volume values (95% CI of 0.53 to 3.7%) when breaths were delivered with 1-s inspiratory times, with similar performance for breaths delivered with 0.5-s inspiratory times (95% CI of 1.1 to 6.7%) and 2-s inspiratory times (95% CI of -1.1 to 2.3%). Audio cues "Bag faster" (1.84 to 2.03 s), "Bag slower" (0.35 to 0.41 s), and "Leak detected" (43 to 50%) were triggered close to target trigger values (2.00 s, 0.50 s, and 50%, respectively) across varying tidal volumes.
Conclusions:
The BENGI achieved its proposed goals of accurately measuring and reporting tidal volumes delivered through BVM systems, providing immediate feedback on the quality of respiratory performance through audio and visual cues. The BENGI has the potential to reduce manual ventilation-induced lung injury and improve patient outcomes by providing accurate feedback on ventilatory parameters.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...