Summary
Compromised cochlear microphonic (CM) amplitude during cochlear implant (CI) surgery indicates poorer postoperative hearing preservation. Preoperative hearing levels influence the ability to record CM signals intraoperatively.
Area of Science:
- Otoacoustic emissions and electrophysiology
- Neurosurgery and audiology
- Medical device monitoring
Background:
- Electrocochleography (ECochG) is an emerging tool for monitoring cochlear function during cochlear implant (CI) surgery.
- Intraoperative ECochG can be recorded from electrodes on the implant array.
- ECochG amplitude changes during electrode insertion may reflect cochlear status.
Purpose of the Study:
- To determine if a drop in ECochG amplitude (compromise) during CI electrode insertion correlates with poorer postoperative acoustic hearing.
- To investigate the influence of preoperative hearing levels on the ability to record ECochG signals.
- To test hypotheses linking CM amplitude compromise to hearing deterioration at different postoperative intervals and preoperative hearing levels to CM recording feasibility.
Main Methods:
- Prospective, multi-site, observational study with 77 adult CI candidates.
- ECochG recorded during and after electrode array implantation.
- Preoperative and postoperative pure-tone audiometry at 4-6 weeks (FUV1) and 3 months (FUV2).
Main Results:
- CM compromise during insertion was associated with significantly poorer hearing preservation at FUV1 (median hearing loss 26.9 dB vs. 12.6 dB, p < 0.014).
- A measurable CM (>1 µV) during implantation depended on preoperative low-frequency hearing thresholds (≤80 dB HL).
- High-frequency preoperative thresholds also correlated with the presence of a measurable CM.
Conclusions:
- Intraoperative CM amplitude drops during CI electrode insertion are linked to worse postoperative hearing preservation.
- Preoperative low-frequency hearing levels below 80 dB HL are important for obtaining reliable intraoperative CM signals.
- ECochG is a practical tool for monitoring cochlear function during CI surgery, with implications for surgical technique and patient outcomes.