Related Experiment Video
Updated: Aug 20, 2025

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Hybrid Closed-Loop Insulin Pump Technology Can Be Safely Used in the Inpatient Setting
Melissa N Pelkey1, Mary E Boyle2, Ashley Long3
1Mayo Clinic Alix School of Medicine, Scottsdale, Arizona.
Background:
Hybrid closed-loop (HCL) systems, also known as automated insulin delivery systems, are a rapidly growing technology in diabetes management. Because more patients are using these systems in the outpatient setting, it is important to also assess inpatient safety to determine whether HCL use can be continued when those patients become hospitalized.
Methods:
The records of patients using HCL technology on admission to our hospital between June 1, 2020, and June 30, 2021, were analyzed.
Results:
The final analysis included 71 patients divided into 3 categories based on their pump use as an inpatient: (1) HCL users; (2) manual pump users; and (3) pump removed. All cohorts were similar in age, sex, race, hemoglobin A1C at admission, and in Medicare Severity Diagnosis Related Group. Pairwise comparisons indicated that patient-stay mean glucose levels, frequency of patient-specific hyperglycemic measurements, and frequency of hypoglycemic events were similar between all groups. No adverse events, particularly occurrences of diabetic ketoacidosis, pump site complications or infection, or equipment malfunction, were reported.
Conclusion:
This preliminary case series review indicates that continued use of HCL technology in the hospital is safe. Moreover, glycemic control in HCL users was comparable with that in those using insulin pump with manual settings and those converted to basal-bolus insulin therapy.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Parentral Nutrition: Centeral and Peripheral Parental Nutrition
PN can be administered through two primary routes:
1. Central Parenteral Nutrition (CPN):
CPN involves delivering a high concentration of nutrients through a large vein. This is typically achieved using a Peripherally Inserted Central Catheter (PICC) or,...
Open and closed-loop control systems
An open-loop control system operates without feedback from the output. It consists of two primary elements: the controller and the controlled process. The controller receives an input signal...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...