Related Experiment Video
Updated: Aug 20, 2025

Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Evaluation of Over-the-Counter Portable Oxygen Concentrators Utilizing a Metabolic Simulator
Richard Casaburi1, Michael Hess2, Janos Porszasz3
1Rehabilitation Clinical Trials Center, Lundquist Institute for Biomedical Innovation at Harbor-UCLA Medical Center, Torrance, California. casaburi@ucla.edu.
Background:
Supplemental oxygen is designed to raise alveolar PO to facilitate diffusion into arterial blood. Oxygen is generally delivered by nasal cannula either by continuous or pulsatile flow. Battery-powered portable oxygen concentrators (POCs) facilitate ambulation in patients experiencing exertional hypoxemia. In the United States, the Food and Drug Administration (FDA) clears these devices to be sold by physician prescription. Recently, however, lower-cost devices described as POCs have been advertised by online retailers. These devices lack FDA clearance and are obtained over the counter (OTC) without prescription. This study determined whether a selected group of OTC POCs have oxygen delivery characteristics suitable for use by hypoxemic patients.
Methods:
A metabolic simulator, capable of simulating a range of metabolic rates and minute ventilations, determined effects of oxygen supplementation delivered by a variety of devices on alveolar PO . Devices tested included 3 OTC POCs, an FDA-cleared POC, and continuous-flow oxygen from a compressed oxygen cylinder. End-tidal PETO , a surrogate of alveolar PO , was determined at each of each device's flow settings at 3 metabolic rates.
Results:
Continuous-flow tank oxygen yielded a linear PETO increase as flow increased, with progressively lower slope of increase for higher metabolic rate. The prescription POC device yielded similar PETO elevations, though with somewhat smaller elevations in pulse-dose operation. One OTC POC was only technically portable (no on-board battery); it provided only modest PETO elevation that failed to increase as flow setting was incremented. A second OTC POC produced only minimal PETO elevation. A third OTC POC, a pulsed-dose device, produced meaningful PETO increases, though not as great as the prescription device.
Conclusions:
Only one of 3 OTC POCs tested was potentially of use by patients requiring ambulatory oxygen. Physicians and respiratory therapists should inform patients requiring portable oxygen that OTC devices may not meet their oxygenation requirements.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...