Related Experiment Video
Updated: Aug 20, 2025

A Standardized Approach to Extra-Oral and Intra-Oral Digital Photography
Published on: July 22, 2022
Improvements in image quality after optimization in digital intraoral radiographs
Background:
Digital intraoral radiographic exposures are optimized largely on the basis of subjective assessment of diagnostic image quality. This study presents an objective approach to optimize radiographic exposure settings for digital intraoral radiographic systems.
Methods:
Seven size 2 digital intraoral systems were assessed for image quality and determination of optimal exposure following the protocol specified in American National Standard Institute/American Dental Association Standard No. 1094: Quality Assurance for Digital Intra-Oral Radiographic Systems. A ProX radiograph unit (Planmeca) at 63 kVp and 6 mA was used to obtain radiographs of the Dental Digital Quality Assurance phantom. ImageJ software (National Institutes of Health) was used to quantify dynamic range and spatial resolution, and contrast perceptibility was evaluated visually. Optimal exposure is the setting with the maximal contrast perceptibility and spatial resolution while displaying the full dynamic range. After image optimization, a custom phantom consisting of an endodontically prepared tooth was imaged to evaluate the file position relative to the apex for each system. Differences in distances between file position relative to the root apex at the optimal exposure as well as 1 increment above and below were measured.
Results:
Radiographic images obtained at the optimal exposure yielded better visualization and more accurate measurements of the file tip relative to the apex.
Conclusions:
Optimizing radiographic exposures improves image quality and accuracy in clinical decisions.
Practical Implications:
Improvement in image quality and better accuracy in actual distance of the endodontic file to the radiographic apex coupled with complete cleaning, shaping, and obturation of the canal should lead to better endodontic treatment outcomes.
Related Concept Videos
X-ray Imaging
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...