Related Experiment Video
Updated: Aug 20, 2025

Delivery of In Vivo Acute Intermittent Hypoxia in Neonatal Rodents to Prime Subventricular Zone-derived Neural Progenitor Cell Cultures
Published on: November 2, 2015
Closed-loop automated oxygen control in ventilated infants born at or near term: A crossover trial
Ourania Kaltsogianni1, Theodore Dassios1,2, Rebecca Lee1
1Women and Children's Health, School of Life Course Sciences, Faculty of Life Sciences and Medicine, King's College London, London, UK.
Aim:
To determine if the use of closed-loop automated oxygen control (CLAC) reduced the incidence and duration of hypoxemic episodes (SpO2 < 92%) in ventilated infants born at or above 34 weeks of gestation.
Methods:
Infants were studied on two consecutive days for 6 h each day. They were randomised to receive standard care (manual oxygen control) or standard care with a CLAC system (automated oxygen control) first.
Results:
Sixteen infants with a median (IQR) gestational age of 37.4 (36.6-38.8) weeks were studied at a median (IQR) postmenstrual age of 38.8 (37.4-39.8) weeks. During the automated oxygen control period, infants spent less time in hypoxemia (SpO2 < 92%) (p = 0.033), episodes of desaturation were shorter (p = 0.001), the time spent within target SpO2 range (92%-96%) was increased (p = 0.001), and the FiO2 delivery was lower (p = 0.018). The time spent in hyperoxemia (SpO2 > 96%) was reduced during automated oxygen control (p = 0.011), the episodes of hyperoxemia were of shorter duration (p = 0.008) and fewer manual adjustments were made to the FiO2 (p = 0.005).
Conclusions:
Closed-loop automated oxygen control in ventilated infants born at or near term was associated with a reduction in the incidence and duration of hypoxemic episodes with more time spent in the target oxygen range.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...