A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Noninvasive Technique to Monitor the Pressure under a Cast: A Mobile Application-Friendly Bluetooth Pressure Sensor
Mahir Mahirogullari1, Serkan Surucu2, Mehmet Halis Cerci1
1Şişli Memorial Hospital, Piyalepaşa Blv. 34385, Şişli, İstanbul, Turkey.
Aim:
The purpose of this study was to design a sensor that could monitor the skin-cast contact surface pressure (SCCSP) of a limb under a cast and inform the user via a mobile application when the pressure increases.
Methods:
In this experimental study, an infant sphygmomanometer cuff was initially placed on the forearm of 10 volunteers. A pressure sensor with a Bluetooth chip was then placed on the volar aspect of the forearm. Short arm plaster was applied with synthetic cast material. The SCCSP under the plaster was measured by the sensor and the measured values were transmitted to a mobile application via a Bluetooth chip. The mobile application processed the data from the chip and converted it to mmHg.
Results:
Intracompartmental pressure (ICP) values were categorized as 0, 10, 20, 30, 40, 50, 60, and 75 mmHg. The highest SCCSP was 75 mmHg CP, while the lowest was 0 mmHg CP. The correlation coefficient of the mean pressure values was 0.993 (p ≤ 0.001) (SD 0.002, range 0.989-0.997), and there was a significant relationship between ICP and SCCSP values (p ≤ 0.05).
Conclusion:
We can monitor SCCSP, detect limb swelling, and notify the user via a mobile application by using Bluetooth pressure sensors.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
06:27Wireless Telemetry Device Implantation in a Fontan Ovine Model for Continuous and Long-Term Hemodynamic Monitoring
Published on: May 2, 2025
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure