Blood pressure measurement in the elderly with atrial fibrillation: an observational study comparing different
Shaojun Mai1, Hailan Zhu1, Meijun Li1
1Department of Cardiology, Shunde Hospital, Southern Medical University, Guangdong, China.
Background:
Atrial fibrillation (AF) has affected millions of adults in the world. It is important to monitor and manage blood pressure (BP) in AF patients. The accuracy of BP monitoring in AF patients with noninvasive methods remains questionable, however.
Objectives:
To compare the accuracy of different noninvasive BP devices (oscillographic sphygmomanometer and pulse wave device) for BP measurement in elderly patients with AF, with a mercury sphygmomanometer as a reference.
Design:
This study was an observational study.
Methods:
Patients with AF from the inpatient department of cardiology were included from 1 January to 31 December 2020. BP measurements were performed by two trained nurses using a tee junction connection on the cuff to connect three sphygmomanometers. The Bland-Altman plot analysis was conducted to compare the agreement of BP measurements. We also compared the agreement of BP measurements through metrics such as accuracy, bias, and precision.
Results:
A total of 202 patients (54.5% female) were included. The Bland-Altman plot analysis showed that the lower and upper limits of agreement (LoAs) of pulse wave/reference were similar to the predefined acceptable clinical limits (10/5 mmHg). The bias and precision in both systolic and diastolic BP were significantly less in pulse wave/reference (a bias of 1.8 and 0.77 mmHg and a precision of 5.20 and 4.66 mmHg, respectively), with corresponding higher accuracy readings (98.51% for P10 in systolic BP and 85.64% for P5 in diastolic BP).
Conclusion:
A novel noninvasive sphygmomanometer - pulse wave device has a good concordance with a mercury sphygmomanometer in BP monitoring, and may be applicable to perform BP measurements in the elderly with AF.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure