Related Experiment Video
Updated: Aug 20, 2025

Evaluation of Oxidative Stress in Biological Samples Using the Thiobarbituric Acid Reactive Substances Assay
Published on: May 12, 2020
Use of a Vanadate Oxidation Conjugated Bilirubin Assay to Reduce Test Cancellations Resulting from Hemolyzed
Benjamin A Kaumeyer1, Melissa Y Tjota1, Kyle Parker1
1Department of Pathology, University of Chicago, Chicago, IL, USA.
Objectives:
We sought to replace the highly hemolysis-susceptible diazo conjugated bilirubin (Bc) assay with the more robust vanadate oxidation method and determine its impact on test cancellation in the pediatric population.
Methods:
Analytical validation of the Randox vanadate assay and comparison with the Roche diazo method were performed. The frequency of pediatric sample cancellation because of hemolysis was compared between the diazo and vanadate methods by retrospective analysis of clinical test data.
Results:
The vanadate assay demonstrated no clinically significant interference from hemolysis up to a hemolysis index of 1,300 (approximately 13 g/L hemoglobin). There was a strong correlation with the diazo method (r2 = 0.97) but with a positive slope bias of 1.27. Implementing the vanadate method resulted in a significantly lower proportion of pediatric samples cancelled because of hemolysis compared with the diazo method (0.6% of 688 patients vs 30.6% of 10,464 patients, respectively; P < .001), with a 0.6% (n = 513) vs 43.2% (n = 6,464) reduction in test cancellations (P < .001) for children younger than 6 months of age.
Conclusions:
The vanadate method showed robust performance against hemolysis. Its implementation resulted in a significant decrease in pediatric tests cancelled because of hemolysis.