Signal Acquisition, Score Interpretation, and Economics of a Non-Invasive Point-of-Care Test for Coronary Artery Disease
Published on: August 9, 2024
Impact of Wire Sensor Location on Coronary Physiological Measurement
Amir Lotfi1, Ali Haider1, Abdullah Pervaiz1
1From the Department of Cardiology, University of Massachusetts Chan Medical School-Baystate, Springfield, MA.
Background:
The location of the wire sensor to measure fractional flow reserve (FFR) and diastolic pressure ratio (dPR) has not been systematically studied. Therefore, we hypothesize that the coronary physiological measurements will vary with the location of the sensor.
Methods:
Fifty-four patients were screened, and 30 consecutive patients were enrolled. The OptoWire 2 or 3 generation fiberoptic pressure wire was used to assess whole cycle pressure distal/pressure aorta, dPR, and FFR. Our primary goal is to test if those measurements vary with the wire sensor placed at 10 mm (proximal), 35-45 mm (mid), and greater than or equal to 60-70 mm (distal) distal to the target lesion, respectively. We used a multilevel linear regression approach.
Results:
Of 30 patients enrolled, 23 (76.6%) were males, mean age was 64.7 years (± 11.0 years), and mean stenosis was 61.6% (±13.4%). Adjusting for age, gender, and severity of stenosis, results showed that for all 3 measures (whole cycle pressure distal/pressure aorta, dPR, and FFR), pressure decreased in a linear fashion the further the sensor was from the target lesion ( P < 0.001). Further, pairwise comparisons of the measurements at adjacent locations similarly showed significant declines in pressure ( P < 0.001).
Conclusions:
This is the first study to demonstrate that the location of the pressure wire can impact the results of both resting and hyperemic pressures, which can cause a false-negative result. This is especially important where the values are near the cutoff.
More Related Videos
07:53Standardized Rat Coronary Ring Preparation and Real-Time Recording of Dynamic Tension Changes Along Vessel Diameter
Published on: June 16, 2022
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Special considerations while measuring pulse
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Location and Orientation of the Heart