A Recirculation System to Reduce the Consumption of Oxygen During CPAP
Andrea Coppadoro1, Luca Paratico2, Giacomo Bellani3
1Department of Anesthesia and Intensive Care, San Gerardo Hospital, ASST Monza, Monza, Italy.
Background:
During continuous flow CPAP for noninvasive respiratory support, a high flow (eg, 60-90 L/min) of gas with FIO titratable up to 1.0 is provided within a helmet or face mask, while a PEEP valve maintains the set pressure. A large amount of oxygen is wasted, whereas only a minimal amount is consumed. We describe a recirculation circuit designed to reuse the exhaust gas and save oxygen.
Methods:
A standard Venturi-based continuous flow system delivering a flow of ≥ 60 L/min to a helmet, and a modified system designed to recirculate oxygen were tested on the bench during simulated active breathing. The proposed system recirculates the oxygen-enriched gas escaping from the PEEP valve; after CO2 removal, the gas is entrained by the Venturi flow generator and redirected to the helmet. We compared oxygen consumption, pneumatic performance, and gas conditioning of the standard and the recirculation systems.
Results:
The recirculation system reduced the oxygen consumption up to 80% as compared to the standard system. Oxygen sparing increased with increasing FIO and total flow delivered to the helmet. Exhaled CO2 was efficiently removed by a single soda lime canister for about 10 h. Pressure swings during a respiratory cycle slightly worsened when using the recirculation system as compared to the standard one (from 3.5-4.4 cm H2O). At FIO 1.0, humidity was 6.1 mg/L and 17.3 mg/L with the standard and the recirculation system, respectively.
Conclusions:
The recirculation system allowed a 80% reduction of oxygen consumption during simulated helmet CPAP therapy, whereas CO2 removal was effective for > 10 h. Recirculation minimally affected pneumatic performance of the CPAP continuous flow system, while improving gas conditioning as compared to the standard system.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Cardiopulmonary Resuscitation II: ACLS Airway Management