Related Experiment Video
Updated: Aug 20, 2025

Cone-Enriched Cultures from the Retina of Chicken Embryos to Study Rod to Cone Cellular Interactions
Published on: March 20, 2021
Tablet-based 'ON/OFF' pathway test can distinguish between rod- and cone-dominated diseases
Amithavikram R Hathibelagal1,2, Phuntsok Bhutia1,2, Mritunjoy Das1,2
1Brien Holden Institute of Optometry and Vision Sciences, L V Prasad Eye Institute, Hyderabad, India.
Introduction:
The photopic ON pathway defect is associated with nocturnal vision loss. However, the measurement of ON function to detect a rod-dominated disease (rods affected more than cones) has not been explored. We evaluated whether the psychophysical evaluation of ON/OFF pathways can be used to distinguish cone-dominated from rod-dominated diseases.
Methods:
Thirty-seven patients with inherited retinal diseases were tested using the 'EyeSpeed' [iOS application] on an iPad. The test displayed a random number (1-3) of light or dark targets on a black-and-white noise background. Participants responded on a touch screen indicating the correct number of targets displayed. The outcome variables-reaction time, accuracy and performance index (speed [1/reaction time] * accuracy) to both light and dark targets were assessed for diagnostic ability using standard receiver-operating characteristic (ROC) analysis.
Results:
Mean ± standard deviation age and visual acuity for the cone- and rod-dominated groups were 25.15 ± 11.74 years, 0.80 ± 0.25 logMAR and 28.3 ± 14.29 years, 0.48 ± 0.26 logMAR, respectively. The median reaction time to light targets in rod-dominated disease [interquartile range] was 5.28 s [3.17], significantly greater than for patients with cone-dominated disease (2.07 s [0.93]; Mann-Whitney U test, p < 0.001). Amongst all of the outcome variables evaluated, the reaction time to light targets (criterion of ≥2.98 s) exhibited the highest area under the ROC curve (area = 0.89 ± 0.11; p < 0.001), with a sensitivity and specificity of 82.4% and 85% respectively.
Conclusions:
Reaction time to light targets using the ON/OFF pathway paradigm is a valid marker to differentiate between rod- and cone-dominated retinal dystrophies. ON pathway function measured using a tablet-based test could act as a supplemental test in the diagnosis of challenging photoreceptor-specific inherited retinal diseases.
Insights
This study shows that measuring reaction time to light stimuli can help differentiate between rod-dominated and cone-dominated inherited retinal diseases. This tablet-based test offers a new way to diagnose these challenging vision disorders.
Area of Science:
- Ophthalmology
- Neuroscience
- Genetics
Background:
- Photopic ON pathway defects are linked to night vision loss.
- Measuring ON pathway function to identify rod-dominated diseases remains unexplored.
- Differentiating between cone- and rod-dominated retinal diseases is crucial for diagnosis.
Purpose of the Study:
- To evaluate the psychophysical assessment of ON/OFF pathways for distinguishing cone-dominated from rod-dominated retinal diseases.
- To explore the diagnostic utility of a tablet-based visual function test.
Main Methods:
- Thirty-seven patients with inherited retinal diseases were tested using the 'EyeSpeed' iPad application.
- The test involved identifying numbers of light or dark targets on a noise background.
- Reaction time, accuracy, and performance index were analyzed using receiver-operating characteristic (ROC) analysis.
Main Results:
- Patients with rod-dominated disease had significantly longer reaction times to light targets (5.28s) compared to cone-dominated disease (2.07s).
- Reaction time to light targets demonstrated the highest diagnostic ability (Area = 0.89, p < 0.001) among all variables.
- A criterion of ≥2.98s for reaction time to light targets achieved 82.4% sensitivity and 85% specificity.
Conclusions:
- Reaction time to light targets, assessed via an ON/OFF pathway paradigm, is a valid marker for differentiating rod- from cone-dominated retinal dystrophies.
- Tablet-based ON pathway function testing can serve as a supplementary diagnostic tool for photoreceptor-specific inherited retinal diseases.