A Multicenter Study on Observed Discrepancies Between Vendor-Stated and PET-Measured 90Y Activities for Both Glass
Silvano Gnesin1, Justin K Mikell2, Maurizio Conti3
1Institute of Radiation Physics, Lausanne University Hospital, University of Lausanne, Lausanne, Switzerland.
Abstract:
Dosimetry-guided treatment planning in selective internal radiation therapy relies on accurate and reproducible measurement of administered activity. This 4-center, 5-PET-device study compared the manufacturer-declared 90Y activity in vials with quantitative 90Y PET/CT assessment of the same vials. We compared 90Y PET-measured activity (APET) for 56 90Y-labeled glass and 18 90Y-labeled resin microsphere vials with the calibrated activity specified by the manufacturer (AM). Additionally, the same analysis was performed for 4 90Y-chloride vials. The mean APET/AM ratio was 0.79 ± 0.04 (range, 0.71-0.89) for glass microspheres and 1.15 ± 0.06 (range, 1.05-1.25) for resin microspheres. The mean APET/AM ratio for 90Y-chloride vials was 1.00 ± 0.04 (range, 0.96-1.06). Thus, we found an average difference of 46% between glass and resin microsphere activity calibrations, whereas close agreement was found for chloride solutions. We expect that the reported discrepancies will promote further investigations to establish reliable and accurate patient dosimetry and dose-effect assessments.