Related Experiment Video
Updated: Aug 20, 2025

Fluorescence Lifetime Macro Imager for Biomedical Applications
Published on: April 7, 2023
Combining fluorescence lifetime with spectral information in fluorescence lifetime imaging ophthalmoscopy (FLIO)
Martin Hammer1,2, Rowena Simon1, Daniel Meller1
1Department of Ophthalmology, University Hospital Jena, Jena, Germany.
Abstract:
Fluorescence lifetime imaging ophthalmoscopy (FLIO) provides information on fluorescence lifetimes in two spectral channels as well as the peak emission wavelength (PEW) of the fluorescence. Here, we combine these measures in an integral three-dimensional lifetime-PEW metric vector and determine a normal range for this vector from measurements in young healthy subjects. While for these control subjects 97 (±8) % (median (interquartile range)) of all para-macular pixels were covered by this normal vector range, it was 67 (±55) % for the elderly healthy, 38 (±43) % for age-related macular degeneration (AMD)-suspect subjects, and only 6 (±4) % for AMD patients. The vectors were significantly different for retinal pigment epithelium (RPE) lesions in AMD patients from that of non-affected tissue (p < 0.001). Lifetime- PEW plots allowed to identify possibly pathologic fundus areas by fluorescence parameters outside a 95% quantile per subject. In a patient follow-up, changes in fluorescence parameters could be traced in the lifetime-PEW metric, showing their change over disease progression.
Related Concept Videos
Confocal Fluorescence Microscopy
Super-resolution Fluorescence Microscopy
Fluorescence and Phosphorescence: Instrumentation