Summary
The ProGlide closure system effectively achieves hemostasis in percutaneous endovascular aortic repair, demonstrating high success rates and low complication rates even in complex cases requiring re-puncture.
Area of Science:
- Vascular Surgery
- Interventional Cardiology
- Medical Device Technology
Background:
- Percutaneous endovascular aortic repair (EVAR) offers a less invasive approach to treating aortic pathologies.
- Achieving reliable hemostasis at the femoral access site is crucial for the success of EVAR procedures.
- Suture-mediated closure devices aim to provide rapid and secure hemostasis, facilitating percutaneous access.
Purpose of the Study:
- To evaluate the efficacy and technical success of the Perclose ProGlide closure device during percutaneous endovascular aortic repair (EVAR).
- To assess intraoperative and postoperative inguinal and vascular complications associated with the ProGlide device in EVAR procedures.
Main Methods:
- A retrospective analysis of patients undergoing EVAR via percutaneous femoral access between April 2017 and June 2021.
- The study analyzed 376 punctures in 263 patients using the Perclose ProGlide vascular closure system.
- Primary endpoints included technical success and efficacy; secondary endpoints focused on perioperative complications.
Main Results:
- The ProGlide system achieved a primary technical success rate of 93.1% and a secondary technical success rate of 94.7%.
- Intraoperative complications requiring femoral exposure occurred in 5.3% of cases, and postoperative complications in 3.5%.
- Larger sheath diameters and percutaneous re-puncture did not significantly increase complication rates or decrease success rates.
Conclusions:
- The ProGlide closure system is a safe and effective method for achieving hemostasis in percutaneous EVAR.
- This closure system is suitable for complex aortic pathologies, including those requiring staged procedures and re-puncture.