Influence of Metal Implants on Quantitative Evaluation of Bone Single-Photon Emission Computed Tomography/Computed
Keisuke Oe1, Feibi Zeng2, Takahiro Niikura1
1Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, Chuo-ku, Kobe 650-0017, Japan.
Abstract:
When visualizing biological activity at nonunion sites by the radioisotopes, gamma rays are more attenuated if metal implants are placed in the bone. However, the effects of various implant types and their placement on gamma ray attenuation in quantitative evaluation remain unknown. To elucidate these effects, we created a phantom that simulated the nonunion of the femur in this study. The count of gamma rays was measured by single-photon emission computed tomography/computed tomography (SPECT/CT) while considering CT-based attenuation correction (CTAC), metal implant placement, type (intramedullary nail or plate), and position. The count differed significantly with and without CTAC and with and without implants (both types) under CTAC. Significantly different counts were observed between the intramedullary nail and plate placed contralaterally to the lesion (i.e., non-lesion side). No significant difference was observed between the intramedullary nail and plate on the lesion side or between plates on the non-lesion and lesion sides. The measured standardized uptake value (SUV) was closer to the true SUV with CTAC than without. Moreover, the count was higher with implants than without. However, even with implants, it was lower than the actual count, indicating the absence of overcorrection. Implant type and position do not seem to influence the count.
More Related Videos
07:48High Spatial Resolution Chemical Imaging of Implant-Associated Infections with X-ray Excited Luminescence Chemical Imaging Through Tissue
Published on: September 30, 2022
06:31Quantitative [18F]-Naf-PET-MRI Analysis for the Evaluation of Dynamic Bone Turnover in a Patient with Facetogenic Low Back Pain
Published on: August 8, 2019