Comparability of childhood blood pressure measurements with two different devices
Lukas Streese1, Christoph Hauser1, Henner Hanssen1
1Department of Sport, Exercise and Health, Medical Faculty, University of Basel, Basel, Switzerland.
Introduction:
Higher blood pressure levels are associated with vascular dysfunction as early as childhood. Here, we aim to compare two blood pressure devices for use in childhood populations within a school setting to screen for elevated blood pressure in children.
Methods:
Systolic and diastolic blood pressure, mean arteriolar pressure (MAP) as well as heart rate (HR) were measured with the validated Oscillomate 9002 and Mindray VS-900 in 82 randomly selected children between 10 and 12 years taking part in the EXAMIN YOUTH study in 2020. Measurements were performed after 5 min at rest in a sitting position in the school environment.
Results:
Both devices showed strong correlations for systolic (r = 0.62, t[80] = 7, p < 0.001) and diastolic blood pressure (r = 0.73, t[80] = 9.52, p < 0.001), MAP (r = 0.75, t[80] = 10.22, p < 0.001) and HR (r = 0.89, t[79] = 17.77, p < 0.001). There was no evidence for a statistically significant difference of both devices for systolic (110 ± 9 mmHg vs. 111 ± 9 mmHg, p = 0.574), diastolic blood pressure (67 ± 9 mmHg vs. 66 ± 9 mmHg, p = 0.301) as well as MAP (81 ± 8 mmHg vs. 80 ± 8 mmHg, p = 0.400) and HR (83 ± 12 BPM vs. 83 ± 11 BPM, p = 0.994).
Discussion:
Both devices, the older Oscillomate 9002 and the current Mindray VS-900 showed good agreement for the measurement of blood pressure in school children. It therefore appears to be feasible to measure childhood blood pressure with either device or replace one device with another in prospective long-term studies or screening programmes as long as both are validated for use in children.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: