Effect of six different autorefractor designs on the precision and accuracy of refractive error measurement
Abinaya Priya Venkataraman1, Rune Brautaset1, Alberto Domínguez-Vicent1
1Division of Eye and Vision, Department of Clinical Neuroscience, Karolinska Institute, Stockholm, Sweden.
Purpose:
To evaluate the precision of objective refraction measurements with six different autorefractors that have different designs and measurement principles and to compare the objective refraction values with the subjective refraction.
Method:
Objective refraction of 55 participants was measured using six autorefractors with different designs. The instrument features mainly varied in terms of measurement principles, inbuilt fogging, open or closed view, and handheld or stationary designs. Two repeated measurements of objective refraction were performed with each autorefractor. The objective refractions from the six autorefractors were compared with the standard subjective refraction. The repeatability limit and Bland-Altman were used to describe the precision and accuracy of each autorefractor, respectively. The analysis was done using the spherical component of the refraction and the power-vector components, spherical equivalent (M), and cylindrical vectors.
Results:
The repeatability of all autorefractors was within 1.00 and 0.35D for measuring the M and both cylindrical components, respectively. Inbuilt fogging was the common feature of the instruments that showed better repeatability. Compared to subjective refraction, the mean difference for sphere and M was below +0.50D, and it was close to zero for the cylindrical components. The instruments that had inbuilt fogging showed narrower limit of agreement. When combined with fogging, the open field refractors showed better precision and accuracy.
Conclusions:
The inbuilt fogging is the most important feature followed by the open view in determining the precision and accuracy of the autorefractor values.
More Related Videos
08:27Quantification of Oculomotor Responses and Accommodation Through Instrumentation and Analysis Toolboxes
Published on: March 3, 2023
06:19Comparison of Three Clinical Stereoscopic Methods for Measuring Binocular Visual Function During Amblyopic Treatment in Unilateral Amblyopia
Published on: September 27, 2024
Related Concept Videos
Focusing of Light in the Eye
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Uncertainty in Measurement: Accuracy and Precision
Distance Corrections
Accuracy and Precision
Common Leveling Mistakes and Errors