The principles and effectiveness of X-ray scatter correction software for diagnostic X-ray imaging: A scoping review
Mohammad Sayed1, Karen M Knapp2, Jon Fulford3
1Diagnostic Radiology Department, College of Applied Medical Sciences, Najran University, Najran 61441, Saudi Arabia; Department of Medical Imaging, College of Medicine and Health, University of Exeter, St Luke's Campus, Heavitree Road, Exeter EX1 2LU, UK.
Purpose:
An anti-scatter grid is often used in X-ray radiography to reduce the scattered X-rays generated from the patient. However, the presence of a grid means the patient dose subsequently increases. Recently,severalmanufacturers have developedsoftwarethat is capable of correctingfor scattered X-rays withouttheuse ofa conventional grid. This scoping review aims to systematically map the research assessing scattering correction software and to identify any existing knowledge gaps.
Methods:
This scoping review involved conducting a systematic search in PubMed, Scopus, and Web of science to reveal studies that were relevant to the research question. Articles published between 01.01.2000 and 31.12.2021 examining X-ray scatter correction software for X-ray imaging were included. A part of the PRISMA model and PICO framework were utilised to establish eligibility criteria. A structured summary table was utilised to extract data from the selected articles.
Results:
In this scoping review, 20 years of literature in X-ray conventional radiography. 11 articles were included in the data synthesis. The study populations of the included studies were varied: patients, image quality phantoms and anatomical phantoms. The clinical applications of X-ray scatter correction software were found to be limited to specific body parts (cervical spine, chest, shoulder, lumbar spine, hip and pelvis). The scatter correction software appears to be effective in terms of image quality and in reducing the radiation dose. However, the conventional grid still provides a higher image quality.
Conclusions:
X-ray scatter correction software can be effective and provides potentialbenefits for some circumstances or clinical scenarios.
More Related Videos
Related Concept Videos
X-ray Imaging
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal...
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...