Analysis of image quality by regulating beta function of BSREM reconstruction algorithm and comparison with
Pooja Dwivedi1,2, Viraj Sawant1,2, Vishal Vajarkar1,2
1Advanced Centre for Treatment Research & Education in Cancer, Tata Memorial Centre, Navi Mumbai.
Background:
The study aimed to evaluate the beta penalization factor of the BSREM reconstruction algorithm on a five-ring BGO-based PET CT system and compared it with conventional reconstructions.
Methods:
Retrospective study involves 30 breast cancer patient data of 18F-fluorodeoxyglucose ( 18 F-FDG) PET CT for reconstruction with OSEM, OSEM + PSF, and BSREM under variable β factors ranging from 200 to 600 in the steps of 50. Liver noise, lesion SUVmax, SBR, and SNR for each reconstruction were calculated. Quantitative parameters of each beta factor of BSREM were compared with OSEM and OSEM + PSF, using the Wilcoxon sign rank test with Bonferroni correction, a value of P < 0.002 was considered statistically significant. Visual scoring by two readers was also evaluated.
Results:
Thirty lesions of mean size 1.91 ± 0.58 cm range (0.7-3.6 cm) were identified. Liver noise and SBR were reduced, whereas SNR was increased with an increasing β value of BSREM. In comparison with OSEM, liver noise was not significantly different from β200 and β250. SNR of OSEM was significantly lower than any other β factors and SBR of β factor less than 500 was significantly higher than OSEM. In comparison with OSEM + PSF, liver noise was not significantly different from β400 and β350-500 do not show a significant difference in SNR and SBR compared with OSEM + PSF. β350 scored highest under visual scoring with a moderate agreement.
Conclusion:
The study quantitatively indicates the optimum beta range of β250-450 and the qualitative evaluation indicates that β350 is an optimum beta factor of BSREM in breast cancer cases for 18 F-FDG WB-PET CT.
More Related Videos
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography