Prospective imaging comparison of anatomic delineation with rapid kV cone beam CT on a novel ring gantry radiotherapy
Lauren E Henke1, Benjamin W Fischer-Valuck1, Soumon Rudra2
1Department of Radiation Oncology, Washington University School of Medicine, St. Louis, MO, United States.
Introduction:
A kV imager coupled to a novel, ring-gantry radiotherapy system offers improved on-board kV-cone-beam computed tomography (CBCT) acquisition time (17-40 seconds) and image quality, which may improve CT radiotherapy image-guidance and enable online adaptive radiotherapy. We evaluated whether inter-observer contour variability over various anatomic structures was non-inferior using a novel ring gantry kV-CBCT (RG-CBCT) imager as compared to diagnostic-quality simulation CT (simCT).
Materials/Methods:
Seven patients undergoing radiotherapy were imaged with the RG-CBCT system at breath hold (BH) and/or free breathing (FB) for various disease sites on a prospective imaging study. Anatomy was independently contoured by seven radiation oncologists on: 1. SimCT 2. Standard C-arm kV-CBCT (CA-CBCT), and 3. Novel RG-CBCT at FB and BH. Inter-observer contour variability was evaluated by computing simultaneous truth and performance level estimation (STAPLE) consensus contours, then computing average symmetric surface distance (ASSD) and Dice similarity coefficient (DSC) between individual raters and consensus contours for comparison across image types.
Results:
Across 7 patients, 18 organs-at-risk (OARs) were evaluated on 27 image sets. Both BH and FB RG-CBCT were non-inferior to simCT for inter-observer delineation variability across all OARs and patients by ASSD analysis (p < 0.001), whereas CA-CBCT was not (p = 0.923). RG-CBCT (FB and BH) also remained non-inferior for abdomen and breast subsites compared to simCT on ASSD analysis (p < 0.025). On DSC comparison, neither RG-CBCT nor CA-CBCT were non-inferior to simCT for all sites (p > 0.025).
Conclusions:
Inter-observer ability to delineate OARs using novel RG-CBCT images was non-inferior to simCT by the ASSD criterion but not DSC criterion.
More Related Videos
10:48PET and MRI Guided Irradiation of a Glioblastoma Rat Model Using a Micro-irradiator
Published on: December 28, 2017
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...